MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1371/journal.pone.0226561

Медицинская библиотека · PloS one · 2019

Thermodilution vs estimated Fick cardiac output measurement in an elderly cohort of patients: A single-centre experience.

Kresoja Karl-Patrik, Faragli Alessandro, Abawi Dawud, Paul Oliver, Pieske Burkert, Post Heiner, Alogna Alessio

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: Patients referred to the cath-lab are an increasingly elderly population. Thermodilution (TD, gold standard) and the estimated Fick method (eFM) are interchangeably used in the clinical routine to measure cardiac output (CO). However, their correlation in an elderly cohort of cardiac patients has not been tested so far.

METHODS

A single, clinically-indicated right heart catheterization was performed on each patient with CO estimated by eFM and TD in 155 consecutive patients (75.1±6.8 years, 57.7% male) between April 2015 and August 2017. Whole Body Oxygen Consumption (VO2) was assumed by applying the formulas of LaFarge (LaF), Dehmer (De) and Bergstra (Be). CO was indexed to body surface area (Cardiac Index, CI).

RESULTS

CI-TD showed an overall moderate correlation to CI-eFM as assessed by LaF, De or Be (r2 = 0.53, r2 = 0.54, r2 = 0.57, all p < .001, respectively) with large limits of agreement (-0.64 to 1.09, -1.07 to 0.77, -1.38 to 0.53 l/m2/min, respectively). The mean difference of CI between methods was 0.22, -0.15 and -0.42 (all p<0.001 for difference to TD), respectively. A rate of error ≥20% occurred with the equations by LaF, De or Be in 40.6%, 26.5% and 36.1% of patients, respectively. A CI <2.2 l/m2min was present in 42.6% of patients according to TD and in 60.0%, 31.0% and in 16.1% of patients according to eFM by the formulas of LaF, De or Be.

CONCLUSION

Although CI-eFM shows an overall reasonable correlation with CI-TD, the predictive value in a single patient is low. CI-eFM cannot replace CI-TD in elderly patients.

Для цитирования: Kresoja Karl-Patrik, Faragli Alessandro, Abawi Dawud, Paul Oliver, Pieske Burkert, Post Heiner, Alogna Alessio. Thermodilution vs estimated Fick cardiac output measurement in an elderly cohort of patients: A single-centre experience.. PloS one. 2019. DOI: 10.1371/journal.pone.0226561.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление