DOI 10.1016/j.brachy.2022.06.006

Медицинская библиотека · Brachytherapy · 2022

Multi-Institutional Study of End-to-End Dose Delivery Quality Assurance Testing for Image-Guided Brachytherapy Using a Gel Dosimeter.

Tachibana Hidenobu, Watanabe Yusuke, Kurokawa Shogo, Maeyama Takuya, Hiroki Tomoyuki, Ikoma Hideaki, Hirashima Hideaki, Kojima Hironori, Shiinoki Takehiro, Tanimoto Yuuki, Shimizu Hidetoshi, Shishido Hiroki, Oka Yoshitaka, Hirose Taka-Aki, Kinjo Masashi, Morozumi Takuya, Kurooka Masahiko, Suzuki Hidekazu, Saito Tomohiko, Fujita Keiichi, Shirata Ryosuke, Inada Ryuji, Yada Ryuichi, Yamashita Mikiko, Kondo Kazuto, Hanada Takashi, Takenaka Tadashi, Usui Keisuke, Okamoto Hiroyuki, Asakura Hiroshi, Notake Ryoichi, Kojima Toru, Kumazaki Yu, Hatanaka Shogo, Kikumura Riki, Nakajima Masaru, Nakada Ryosei, Suzuki Ryusuke, Mizuno Hideyuki, Kawamura Shinji, Nakamura Mistuhiro, Akimoto Tetsuo

Многоцентровое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To quantify dose delivery errors for high-dose-rate image-guided brachytherapy (HDR-IGBT) using an independent end-to-end dose delivery quality assurance test at multiple institutions. The novelty of our study is that this is the first multi-institutional end-to-end dose delivery study in the world.

MATERIALS AND

METHODS

The postal audit used a polymer gel dosimeter in a cylindrical acrylic container for the afterloading system. Image acquisition using computed tomography, treatment planning, and irradiation were performed at each institution. Dose distribution comparison between the plan and gel measurement was performed. The percentage of pixels satisfying the absolute-dose gamma criterion was reviewed.

RESULTS

Thirty-five institutions participated in this study. The dose uncertainty was 3.6% ± 2.3% (mean ± 1.96σ). The geometric uncertainty with a coverage factor of k = 2 was 3.5 mm. The tolerance level was set to the gamma passing rate of 95% with the agreement criterion of 5% (global)/3 mm, which was determined from the uncertainty estimation. The percentage of pixels satisfying the gamma criterion was 90.4% ± 32.2% (mean ± 1.96σ). Sixty-six percent (23/35) of the institutions passed the verification. Of the institutions that failed the verification, 75% (9/12) had incorrect inputs of the offset between the catheter tip and indexer length in treatment planning and 17% (2/12) had incorrect catheter reconstruction in treatment planning.

CONCLUSIONS

The methodology should be useful for comprehensively checking the accuracy of HDR-IGBT dose delivery and credentialing clinical studies. The results of our study highlight the high risk of large source positional errors while delivering dose for HDR-IGBT in clinical practices.

Для цитирования: Tachibana Hidenobu, Watanabe Yusuke, Kurokawa Shogo, Maeyama Takuya, Hiroki Tomoyuki, Ikoma Hideaki, Hirashima Hideaki, Kojima Hironori, Shiinoki Takehiro, Tanimoto Yuuki, Shimizu Hidetoshi, Shishido Hiroki, Oka Yoshitaka, Hirose Taka-Aki, Kinjo Masashi, Morozumi Takuya, Kurooka Masahiko, Suzuki Hidekazu, Saito Tomohiko, Fujita Keiichi, Shirata Ryosuke, Inada Ryuji, Yada Ryuichi, Yamashita Mikiko, Kondo Kazuto, Hanada Takashi, Takenaka Tadashi, Usui Keisuke, Okamoto Hiroyuki, Asakura Hiroshi, Notake Ryoichi, Kojima Toru, Kumazaki Yu, Hatanaka Shogo, Kikumura Riki, Nakajima Masaru, Nakada Ryosei, Suzuki Ryusuke, Mizuno Hideyuki, Kawamura Shinji, Nakamura Mistuhiro, Akimoto Tetsuo. Multi-Institutional Study of End-to-End Dose Delivery Quality Assurance Testing for Image-Guided Brachytherapy Using a Gel Dosimeter.. Brachytherapy. 2022. DOI: 10.1016/j.brachy.2022.06.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.