Neural response telemetry measures in patients implanted with Nucleus 24.

Guedes Mariana Cardoso, Brito Neto Rubens V, Gomez Maria Valéria S Goffi, Sant'Anna Sandra B Giorgi, Peralta Cristina G Ornelas, Castilho Arthur Menino, Bento Ricardo Ferreira

СтатьяПеревод в очереди

Аннотация

UNLABELLED: Cochlear implantation has been recommended for children under 24 months of age. The use of objective measures is needed to help speech processor programming. The electrically evoked compound potential (EAP), which can be assessed by neural response telemetry (NRT), is one of those objective measures.

AIM: to determine how often the EAP can be recorded by NRT system during surgery and to describe the responses.

STUDY

DESIGN

Clinical with transversal cohort.

MATERIAL AND METHOD: The impedances and NRT were measured in a group of 17 Nucleus 24 implant users. The responses were analyzed and compared to the etiology, hearing loss duration and electrode array position.

RESULTS

The EAP was easily recorded in the apical electrodes and, in otosclerosis and meningitis cases the EAP threshold was higher than in the other etiology cases.

CONCLUSIONS

The NRT can be found in 82% of the cases during surgery. The responses obtained may vary according to etiology and the position of electrodes along the cochlea.

Для цитирования: Guedes Mariana Cardoso, Brito Neto Rubens V, Gomez Maria Valéria S Goffi, Sant'Anna Sandra B Giorgi, Peralta Cristina G Ornelas, Castilho Arthur Menino, Bento Ricardo Ferreira. Neural response telemetry measures in patients implanted with Nucleus 24.. Brazilian journal of otorhinolaryngology. 2006. DOI: 10.1016/s1808-8694(15)31271-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.