DOI 10.1080/14670100.2025.2555115
Comparison of photon-counting CT and conventional CT for cochlear implant preoperative planning.
Аннотация
INTRODUCTION
Accurate cochlear measurements are crucial for personalized cochlear implantation, requiring high-quality imaging for precise cochlear duct length (CDL) assessment. Photon-counting computed tomography (PC-CT) is an advanced imaging technique that enhances image quality while reducing radiation exposure. The aim of this study was to compare PC-CT and conventional CT in cochlear measurements using cochlear implant planning software in patients with a unilateral cochlear implant.
METHOD: This prospective study was conducted at a tertiary referral center. All patients had a previous conventional temporal bone CT before cochlear implantation and a PC-CT of the temporal bone for the post implant evaluation. Image analysis for both PC-CT and CT was performed with the OTOPLAN version 3.0 software manually and independently by three different otologist experts in the field.
RESULTS
Ten cochleae were evaluated. The mean full CDL was 35.81 ± 1.22 mm in CT and 35.99 ± 1.94 mm in PC-CT (P = 0.77) Inter-observer reliability was poor for CT with an intra-correlation coefficient of 0.16 and moderate for PC-CT with an intra-correlation coefficient of 0.71.
CONCLUSION
This study demonstrated the advantages of PC-CT over conventional CT by improving the reliability and reproducibility of CDL measurements in patients.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.