INTRODUCTION

The rapid and accurate detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is required to prevent the spread of COVID-19. This study evaluated the utility of two SARS-CoV-2 antigen detection methods.

METHODS

We evaluated two types of antigen detection methods using immunochromatography (Espline) and quantitative chemiluminescent enzyme immunoassay (Lumipulse). RT-PCR was performed as a standard procedure for COVID-19 diagnosis. Lumipulse and RT-PCR were performed for all 486 nasopharyngeal swabs and 136 saliva samples, and the Espline test was performed for 271 nasopharyngeal swabs and 93 saliva samples.

RESULTS

The sensitivity and specificity of the Espline test were 10/11 and 260/260 (100%), respectively for the nasopharyngeal swabs and 3/9 and 84/84 (100%), respectively for the saliva samples. High sensitivities for both saliva (8/9) and nasopharyngeal swabs (22/24) were observed in the Lumipulse test. The specificities of the Lumipulse test for nasopharyngeal swabs and saliva samples were 460/462 (99.6%) and 123/127 (96.9%), respectively.

CONCLUSION

The Espline test is not effective for saliva samples but is useful for simple and rapid COVID-19 tests using nasopharyngeal swabs because it does not require special devices. The Lumipulse test is a powerful high-throughput tool for COVID-19 diagnosis because it has high detection performance for nasopharyngeal swabs and saliva samples.

Для цитирования: Ishii Toshiaki, Sasaki Masakazu, Yamada Kageto, Kato Daiki, Osuka Hiroyoshi, Aoki Kotaro, Morita Toshisuke, Ishii Yoshikazu, Tateda Kazuhiro. Immunochromatography and chemiluminescent enzyme immunoassay for COVID-19 diagnosis.. Journal of infection and chemotherapy : official journal of the Japan Society of Chemotherapy. 2021. DOI: 10.1016/j.jiac.2021.02.025.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.