Histologic effects of a new device for high-intensity focused ultrasound cyclocoagulation.

Aptel Florent, Charrel Thomas, Palazzi Xavier, Chapelon Jean-Yves, Denis Philippe, Lafon Cyril

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the histologic effects and clinical outcomes of high-intensity focused ultrasound (HIFU) delivered by miniaturized annular transducers for ciliary body coagulation in an animal study.

METHODS

Eighteen eyes of 18 rabbits were insonified with a ring comprising a six-sector transducer, to produce a 12.8-mm-diameter circular lesion. Six sectors were activated in six rabbits (group 1), five sectors in six rabbits (group 2), and four sectors in six rabbits (group 3) at 2 W acoustic power. The rabbits were examined before treatment (day 0) and after treatment on days 1, 7, 15, 21, and 28. Detailed qualitative and semiquantitative histopathologic analyses of the enucleated eyes were performed.

RESULTS

In the treated eyes, intraocular pressure changes ranged from -16.6 mm Hg (-55.3%) at day 28 to -8.9 mm Hg (-29.7%) at day 7 in group 1, from -4.7 mm Hg (-25.5%) at day 28 to -1.4 mm Hg (-7.6%) at day 21 in group 2 and from -7.9 mm Hg (-28.1%) at day 28 to +2.0 mm Hg (+7.1%) at day 7 in group 3. No macroscopic abnormalities were observed in the anterior segment or fundus. Histologic examination showed segmental-to-annular lesions in the ciliary processes, caused mainly by coagulation necrosis, whereas the sclera and lens appeared undamaged. Inflammation was very limited.

CONCLUSIONS

Ultrasonic coagulation of the ciliary body with HIFU delivered via a circular miniaturized transducer seemed to be an effective and well-tolerated method of reducing intraocular pressure in an animal study.

Для цитирования: Aptel Florent, Charrel Thomas, Palazzi Xavier, Chapelon Jean-Yves, Denis Philippe, Lafon Cyril. Histologic effects of a new device for high-intensity focused ultrasound cyclocoagulation.. Investigative ophthalmology & visual science. 2010. DOI: 10.1167/iovs.09-5135.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.