DOI 10.1002/mp.17800

Медицинская библиотека · Medical physics · 2025

Dosimetric impact of positional uncertainties and a robust optimization approach for rectal intensity-modulated brachytherapy.

Morén Björn, Thibodeau-Antonacci Alana, Kalinowski Jonathan, Enger Shirin A

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Intensity-modulated brachytherapy (IMBT) employs rotating high-Z shields during treatment to decrease radiation in certain directions and conform the dose distribution to the target volume. Prototypes for dynamic IMBT have been proposed for prostate, cervical, and rectal cancer.

PURPOSE: We considered two shielded applicators for IMBT rectal cancer treatment and investigated how rotational uncertainties in the shield angle and translational uncertainties in the source position affect plan evaluation criteria.

METHODS

The effect of rotational errors of 3 ∘ $3^\circ$ , 5 ∘ $5^\circ$ and 10 ∘ $10^\circ$ , and translational errors of 1, 2 and 3 mm on evaluation criteria were investigated for shields with 180 ∘ ${\rm 180}^\circ$ and 90 ∘ ${\rm 90}^\circ$ emission windows. Further, a robust optimization approach based on quadratic penalties that includes scenarios with errors was proposed. The extent to which dosimetric effects of positional errors can be mitigated with this model was evaluated compared to a quadratic penalty model without scenarios with errors. A retrospective rectal cancer data set of ten patients was included in this study. Treatment planning was performed using the Monte Carlo-based treatment planning system, RapidBrachyMCTPS.

RESULTS

For the largest investigated rotational error of ± 10 ∘ $\pm 10^\circ$ , the clinical target volume  D 90 ${\rm D}_{90}$ remained, on average, within 5 % $5\%$ of the result without error, while the contralateral healthy rectal wall experienced an increase in the mean D 0.1 c c ${\rm D}_{0.1cc}$ , D 2 c c ${\rm D}_{2cc}$ , and D 50 ${\rm D}_{50}$ of 26 % $26\%$ , 9 % $9\%$ , and 1 % $1\%$ for the 180 ∘ ${\rm 180}^\circ$ shield and of 32%, 9%, and 2% for the 90 ∘ ${\rm 90}^\circ$ shield. For translational errors of ± 2 $\pm 2$  mm, there were increases in dosimetric indices for both the superior (sup) and inferior (inf) dose spill regions. Specifically, for the 180 ∘ ${\rm 180}^\circ$ shield, the D 0.1 c c ${\rm D}_{0.1cc}$ , D 2 c c ${\rm D}_{2cc}$ , and D 50 ${\rm D}_{50}$ increased by 13 % $13\%$ , 11 % $11\%$ , and 10 % $10\%$ , respectively, for the sup region, and by 26 % $26\%$ , 15 % $15\%$ , and 11 % $11\%$ , respectively, for the inf region. Similar results were obtained with the 90 ∘ ${\rm 90}^\circ$ shield. Overall, the robust and traditional models had similar results. However, the number of active dwell positions obtained with the robust model was larger, and the longest dwell time was shorter.

CONCLUSIONS

We have quantified the effect of rotational shield and translational source errors of various magnitudes on evaluation criteria for rectal IMBT. The robust optimization approach was generally not able to mitigate positional errors. However, it resulted in more homogeneous dwell times, which can be beneficial in conventional high-dose-rate brachytherapy to avoid hot spots around specific dwell positions.

Для цитирования: Morén Björn, Thibodeau-Antonacci Alana, Kalinowski Jonathan, Enger Shirin A. Dosimetric impact of positional uncertainties and a robust optimization approach for rectal intensity-modulated brachytherapy.. Medical physics. 2025. DOI: 10.1002/mp.17800.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.