ICD Implantation in infants and small children: the extracardiac technique.

Kriebel Thomas, Ruschewski Wolfgang, Gonzalez y Gonzalez Maria, Walter Katharina, Kroll Johannes, Kampmann Christoph, Heinemann Markus, Schneider Heike, Paul Thomas

Клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

There is no clear methodology for implantation of an internal cardioverter-defibrillator (ICD) in infants and small children. The aim of this study was to assess efficacy and safety of an extracardiac ICD implantation technique in pediatric patients.

PATIENTS AND

METHODS

An extracardiac ICD system was implanted in eight patients (age: 0.3-8 years; body weight: 4-29 kg). Under fluoroscopic guidance a defibrillator lead was tunneled subcutaneously starting from the anterior axillar line along the course of the 6th rib until almost reaching the vertebral column. After a partial inferior sternotomy, bipolar steroid-eluting sensing and pacing leads were sutured to the atrial wall (n = 2) and to the anterior wall of the right ventricle (n = 8). The ICD device was implanted as "active can" in the upper abdomen. Sensing, pacing, and defibrillation thresholds (DFTs) as well as impedances were verified intraoperatively and 3 months later, respectively.

RESULTS

In seven of eight patients, intraoperative DFT between subcutaneous lead and device was <15 J. In the eighth patient ICD implantation was technically not feasible due to a DFT >20 J. During follow-up (mean 14.5 months) appropriate and effective ICD discharges were noted in two patients. DFT remained stable after 3 months in four of six patients retested. A revision was required in one patient due to lead migration and in another patient due to a lead break.

CONCLUSIONS

In infants and small children, extracardiac ICD implantation was technically feasible. Experience and follow-up are still limited. The course of the DFT is unknown, facing further growth of the patients.

Для цитирования: Kriebel Thomas, Ruschewski Wolfgang, Gonzalez y Gonzalez Maria, Walter Katharina, Kroll Johannes, Kampmann Christoph, Heinemann Markus, Schneider Heike, Paul Thomas. ICD Implantation in infants and small children: the extracardiac technique.. Pacing and clinical electrophysiology : PACE. 2006. DOI: 10.1111/j.1540-8159.2006.00542.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.