MedElevator Войти в исследования
← Поиск публикаций

DOI 10.23736/s0022-4707.18.08339-1

Медицинская библиотека · The Journal of sports medicine and physical fitness · 2018

Measuring the validity and reliability of the Apple Watch as a physical activity monitor.

Zhang Peng, Godin Steven D, Owens Matthew V

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

This study aimed to investigate the validity and reliability of the energy expenditure (EE) estimation of Apple Watch® among college students.

METHODS

Thirty college students completed two sets of three 10-minute treadmill walking and running trials while wearing three Apple watches and being connected to indirect calorimetry. The walking trials were at speeds of 54, 80, and 107 m·min-1 while the running trials were at 134, 161, 188 m·min-1. Energy expenditure comparisons were made using Two-way ANOVA with repeated measures. Reliability was analyzed by Intraclass Correlation.

RESULTS

There was no significant device x speed interactions (F (15, 696)=1.113, P=0.341) between the indirect calorimetry (criterion) and Apple Watch®. The lowest Inter-Class Correlation (ICC) scores were 0.49 (95%CI) at 54 while the highest were 0.72 (95%CI) at 107 and 134 m·min-1.

CONCLUSIONS

Apple Watch® demonstrated a low to moderate validity and reliability on measuring EE.

Для цитирования: Zhang Peng, Godin Steven D, Owens Matthew V. Measuring the validity and reliability of the Apple Watch as a physical activity monitor.. The Journal of sports medicine and physical fitness. 2018. DOI: 10.23736/s0022-4707.18.08339-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.