Clinical transition to model-based dose calculation algorithm: A retrospective analysis of high-dose-rate tandem and ring brachytherapy of the cervix.

Jacob Dayee, Lamberto Melissa, DeSouza Lawrence Lana, Mourtada Firas

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To retrospectively compare clinical dosimetry of CT-based tandem-ring treatment plans using a model-based dose calculation algorithm (MBDCA) with the standard TG-43-based dose formalism.

METHODS AND MATERIALS: A cohort of 10 cervical cancer cohorts treated using the tandem and ring high-dose-rate applicators were evaluated. The original treatment plans were created using the department CT-based volume optimization clinical standards. All plans originally calculated with TG-43 dose calculation formalism were recalculated using the MBDCA algorithm. The gross target volume and organs at risk (OARs) were contoured on each data set along with significant heterogeneities like air in cavity and high-density plastic tandem and ring components. The patient tissue was modeled as homogenous liquid water. D90, D95, and D100 for gross target volume, D0.1cm3, D1.0cm3, and D2.0cm3 for bladder, rectum, and sigmoid were extracted from dose-volume histograms for TG-43 and MBDCA calculated plans. Mean absolute difference ± 2σ in the above metrics was calculated for each plan.

RESULTS

Using the manual applicator contouring method, MBDCA plans (n = 10) showed 2.1 ± 1.1% reduction in dose to Point A average, 2.6 ± 0.9% reduction in Target D90 dose, and 2.1 ± 0.3% dose reduction to OARs. Results from plans using vendor supplied solid applicator models (n = 5) showed 2.2 ± 1.10% reduction in dose to Point A average, 2.7 ± 0.2% reduction in Target D90 dose, and 2.7 ± 1.0% dose reduction on average to OARs.

CONCLUSION

For unshielded plastic gynecologic applicators, minimal dosimetric changes (<5%) were found using MBDCA relative to standard TG-43. Use of solid applicator model is more efficient than manual applicator contouring and also yielded similar MBDCA dosimetric results. Currently, TG-186 dose calculations should be reported along TG-43 until we obtain studies with larger cohorts to fully realize the potential of MBDCA dosimetry.

Для цитирования: Jacob Dayee, Lamberto Melissa, DeSouza Lawrence Lana, Mourtada Firas. Clinical transition to model-based dose calculation algorithm: A retrospective analysis of high-dose-rate tandem and ring brachytherapy of the cervix.. Brachytherapy. 2017. DOI: 10.1016/j.brachy.2017.02.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.