Comparison of the repeatability of GABA-edited magnetic resonance spectroscopy with and without macromolecule suppression.

Mikkelsen Mark, Singh Krish D, Sumner Petroc, Evans C John

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The inhibitory neurotransmitter γ-aminobutyric acid (GABA) can be measured in vivo using edited magnetic resonance spectroscopy (MRS), but quantification suffers from contamination by macromolecules (MM). It is possible to suppress this contamination using symmetric editing, but this procedure potentially compromises reliability of the GABA measurement. The aim of this study was to compare the repeatability of GABA-edited MRS with and without MM suppression.

METHODS

GABA' (non-MM contaminated) and GABA'+MM (MM-contaminated) concentration was measured in the occipital lobe (OCC) and anterior cingulate (AC) using symmetric and standard editing (n = 15). Each method was performed twice in each region.

RESULTS

Within-participant coefficients of variation for each technique were 4.0% (GABA'+MM) and 8.6% (GABA') in the OCC and 14.8% (GABA'+MM) and 12.6% (GABA') in the AC. Intraclass correlation coefficients were better for the suppression method than standard editing in both the OCC (0.72 versus 0.67) and AC (0.41 versus 0.16). These findings were replicated in the OCC of a second cohort (n = 15).

CONCLUSION

Symmetric suppression is shown to be comparable in repeatability to standard GABA-editing. Measuring a purer quantification of GABA becomes increasingly important as more research is conducted on links between GABA concentration, pathology and healthy behavior.

Для цитирования: Mikkelsen Mark, Singh Krish D, Sumner Petroc, Evans C John. Comparison of the repeatability of GABA-edited magnetic resonance spectroscopy with and without macromolecule suppression.. Magnetic resonance in medicine. 2015. DOI: 10.1002/mrm.25699.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.