DOI 10.1007/s00247-025-06209-2

Медицинская библиотека · Pediatric radiology · 2025

Photon-counting CT yields superior abdominopelvic image quality at lower radiation and iodinated contrast doses.

Schwartz Fides Regina, Bache Steve, Lee Rachel, Maxfield Charles M, Fadell Michael F, Gaca Ana M, Samei Ehsan, Frush Donald P, Cao Joseph Y

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Photon-counting detector (PCD) computed tomography (CT) has been shown to provide better image quality at lower radiation and intravenous contrast doses than energy-integrating detector (EID) CT in adult patients. There is limited data on these benefits for the pediatric population especially for abdominopelvic CT examinations.

OBJECTIVE

This study examines a reduced weight-based iodinated contrast dosing strategy in pediatric abdominopelvic CT on a PCD-CT system compared to standard dosing protocols on EID-CT using 1 mL/kg and 2 mL/kg, respectively. Image quality is assessed using both quantitative and qualitative measures. We also compare the radiation dose profile between the two PCD-CT and EID-CT cohorts.

MATERIALS AND

METHODS

This HIPAA-compliant, IRB-approved, retrospective study included pediatric patients (≤18 years of age) who underwent contrast-enhanced CT examinations of the abdomen and pelvis for routine clinical care (01/2022 - 01/2023) in the portal-venous phase on a PCD-CT (NAEOTOM Alpha; Siemens Healthineers). Inclusion criteria included a similar prior examination within 12-months on a dual-source EID-CT scanner from the same vendor. All PCD-CT and EID-CT scans were acquired using weight-based dosing for intravenous contrast media, 1 mL/kg and 2 mL/kg, respectively, based on institutional protocols. Contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) were measured in the aorta, portal vein, liver parenchyma, and skeletal muscle. Three pediatric radiologists qualitatively evaluated each scan for overall image quality, noise, and contrast on a scale of 0-100. Confidence in small structure detection (common bile duct) was also rated on a scale of 0-3. Radiation doses (size-specific dose estimate (SSDE)) were calculated. Statistical analysis included paired t-tests and a mixed linear effects model to account for patient age, sex, and X-ray tube voltage. A P<0.05 indicated statistical significance.

RESULTS

A total of 49 patients were included (24 female; mean [SD] age 9.9 [6.3] years, range 0.6-18 years). Compared to EID-CT, PCD-CT had a higher mean SNR in the portal vein (23.4 [SD=9.3] vs 17.2 [SD=7.4], P<0.001), aorta (23.4 [SD=11.6] vs 17.7 [10.1], P=0.017), hepatic parenchyma (15.2 [SD=5.6] vs 13.2 [5.1], P=0.016), and skeletal muscle (5.7 [SD=3.1] vs 4.5 [SD=3.1], P=0.01). Compared to EID-CT, PCD-CT also had a higher mean CNR in the portal vein (27.5 [SD=9.6] vs 22.1 [SD=21.1], P=0.003), aorta (27.3 [SD=9.6] vs 22.3 [SD=11.8], P=0.004), hepatic parenchyma (20 [SD=6.9] vs 16.9 [SD=8.5], P=0.013), and skeletal muscle (14.6 [4.9] vs 12.1 [5.6], P=0.008). Overall image quality, image noise, and small structure detection confidence scores were higher on PCD-CT than EID-CT (P=0.037, P<0.001, and P=0.006, respectively). Mean SSDE for PCD-CT was lower than EID-CT (9.1 mGy [SD=4.3] vs 11 mGy [5.9], P=0.012).

CONCLUSION

Compared with EID-CT, contrast-enhanced pediatric abdominopelvic CT offers improved subjective and objective image quality, even at lower radiation doses and reduced intravenous contrast medium volumes.

Для цитирования: Schwartz Fides Regina, Bache Steve, Lee Rachel, Maxfield Charles M, Fadell Michael F, Gaca Ana M, Samei Ehsan, Frush Donald P, Cao Joseph Y. Photon-counting CT yields superior abdominopelvic image quality at lower radiation and iodinated contrast doses.. Pediatric radiology. 2025. DOI: 10.1007/s00247-025-06209-2.