MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/mao.0b013e31822e9513

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2011

Treatment of the atretic ear with round window vibrant soundbridge implantation in infants and children: electrocochleography and audiologic outcomes.

Mandalà Marco, Colletti Liliana, Colletti Vittorio

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the long-term outcomes of the first 5 infants and 9 children with congenital aural atresia (CAA) who had undergone hearing rehabilitation using the MED-EL Vibrant Soundbridge with intraoperative assistance of electrocochleography (ECoG) for optimal fitting of the floating mass transducer (FMT) on the round window (RW) membrane.

STUDY

DESIGN

Tertiary referral medical center; retrospective case series.

PATIENTS: Infants and children ranging in age from 2 months to 16 years with a moderate-to-severe conductive or mixed hearing loss with CAA. For comparison, the study population was divided into 2 groups: older children (≥5 yr of age; 5 patients) and younger children/infants (<5 yr of age; 9 subjects) who were submitted to different audiologic tests appropriate for their age and general condition.

INTERVENTION: RW implantation.

MAIN OUTCOME MEASURES: Compound action potential threshold and amplitude were assessed as a function of different methods for stabilizing the FMT on the RW. Pure tone audiogram at 0.5, 1, 2, and 4 kHz, free-field speech testing (older children), bone conduction and free-field auditory brainstem response (ABR; younger children and infants), intraoperative and postoperative complications, and FMT displacement or extrusion rate.

RESULTS

Statistically significant differences were observed with ECoG recordings between pre- and post-FMT-RW membrane optimization with fascia and cartilage (p < 0.001). Significant improvements were observed in speech perception and pure-tone and ABR threshold, immediately after surgery and at follow-up intervals (12-65 mo) in children and infants (p < 0.01). No complications or instances of device extrusion were observed.

CONCLUSION

Infants and children with moderate-to-severe conductive or mixed hearing loss, not suitable or unwilling to accept Bone-Anchored Hearing Aids and who would not tolerate traditional bone and air conduction hearing aids, obtain substantial benefit with the FMT-RW implantation procedure. Intraoperative ECoG is of significant help in achieving the best FMT-RW fitting.

Для цитирования: Mandalà Marco, Colletti Liliana, Colletti Vittorio. Treatment of the atretic ear with round window vibrant soundbridge implantation in infants and children: electrocochleography and audiologic outcomes.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2011. DOI: 10.1097/mao.0b013e31822e9513.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.