Comparison of patient dose from imaging protocols for dental implant planning using conventional radiography and computed tomography.

Lecomber A R, Yoneyama Y, Lovelock D J, Hosoi T, Adams A M

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To compare the radiation doses from imaging protocols for dental implant planning either using conventional radiography only (dental panoramic radiography (DPR), cephalometry and linear cross-sectional tomography) or involving computed tomography (CT).

METHODS

Organ absorbed doses were measured using a female Rando anthropomorphic phantom loaded with lithium fluoride thermoluminescent dosemeters (TLD). Standard mandibular protocols for dental implant planning were followed using either a conventional dental radiographic unit (PM 2002 CC Planmeca, Helsinki, Finland) or CT scanner (Excel Twin Elscint, Haifa, Israel). Organ absorbed and effective doses were calculated. Effective dose was calculated using two approaches, one based on the ICRP method which excludes the salivary tissue from the remainder organs (designated E(exc)), and the other with its inclusion (E(inc)).

RESULTS

The greatest individual organ doses for any examination were measured in the salivary tissue. E(exc) for panoramic, cephalometric and cross-sectional tomography using DPR was 0.004 mSv, 0.002 mSv and 0.002 mSv, respectively, whereas with CT it was 0.314 mSv. The value of E(inc) calculated using these data was between two and five times E(exc).

CONCLUSIONS

E(inc) greatly increases the apparent radiation burden, especially with high dose procedures. CT techniques can provide excellent images, but at the cost of increased radiation detriment. DPR with a cross-sectional tomography facility may give adequate clinical information at a greatly reduced dose.

Для цитирования: Lecomber A R, Yoneyama Y, Lovelock D J, Hosoi T, Adams A M. Comparison of patient dose from imaging protocols for dental implant planning using conventional radiography and computed tomography.. Dento maxillo facial radiology. 2001. DOI: 10.1038/sj/dmfr/4600627.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.