MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcrs.2007.10.048

Медицинская библиотека · Journal of cataract and refractive surgery · 2008

Acute angle closure after implantable contact lens insertion unresponsive to surgical peripheral iridectomy.

Chan Kenneth C Y, Birchall Wayne, Gray Trevor B, Wells Anthony P

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

We present a case of acute angle closure that occurred after insertion of an implantable contact lens (ICL). The apparent papillary-block angle closure did not resolve after 2 patent iridotomies and a surgical iridectomy, but did respond to pupil dilation (not constriction). Ultrasound biomicroscopy revealed abnormally large and irregular ciliary processes that may have contributed to the unusual behavior of the ICL-iris complex. The condition resolved after the ICL was replaced by one with a smaller haptic diameter. Routine ultrasound biomicroscopic assessment of the ciliary body anatomy preoperatively and ICL haptic positioning postoperatively may identify risk factors that could predispose ICL patients to acute angle closure.

Для цитирования: Chan Kenneth C Y, Birchall Wayne, Gray Trevor B, Wells Anthony P. Acute angle closure after implantable contact lens insertion unresponsive to surgical peripheral iridectomy.. Journal of cataract and refractive surgery. 2008. DOI: 10.1016/j.jcrs.2007.10.048.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.