Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2022
Technical feasibility and clinical evaluation of 4D-MRI guided liver SBRT on the MR-linac.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
van de Lindt T N, Nowee M E, Janssen T, Schneider C, Remeijer P, van Pelt V W J, Betgen A, Jansen E P M, Sonke J J
Аннотация
PURPOSE: Image-guided stereotactic body radiation therapy (SBRT) is an important local treatment for liver metastases. MRI-guidance enables direct tumor visualization, eliminating fiducial marker implantation. The purpose of this study was to test technical feasibility of our 4D-MRI guided liver SBRT workflow. Additionally, intra-fraction target motion and consequent target-coverage were studied.
MATERIALS & METHODS: Patients with liver metastases were included in this sub-study of the prospective UMBRELLA-II clinical trial. Patients received mid-position (midP) SBRT. The daily adapt-to-position workflow included localization, verification and intra-fraction tumor midP monitoring using 4D-MRI. Technical feasibility was established based on persistence of the treatment protocol, treatment time ≤1 h, no geographical miss and no unexpected acute toxicity grade >3. All 4D-MRIs were registered to the planning midP-CT and tumor midP and amplitude were calculated. Additionally, delivered target dose was accumulated incorporating the 4D-MRI intra-fraction tumor motion and evaluated with Monte-Carlo error simulations.
RESULTS: 20 patients with liver metastases were included and treated with 4D-MRI guided SBRT. Feasibility criteria were met in all-but-one patient. No grade ≥3 acute toxicity was observed. Group mean (M), systematic and random midP-drifts were 2.4 mm, 2.6 mm and 3.1 mm in CC-direction. 4D-MRI tumor CC-amplitudes were reduced compared to the simulation 4D-CT (M = -1.9 mm) and decreased during treatment (M = -1.4 mm). Dose accumulation showed adequate target-coverage on a population level.
CONCLUSION: We successfully demonstrated technical feasibility of 4D-MRI guided SBRT in a cohort of 20 patients with liver metastases. However, substantial midposition drifts occurred which stress the need for intra-fraction motion management strategies to further increase the precision of treatment delivery.
MATERIALS & METHODS: Patients with liver metastases were included in this sub-study of the prospective UMBRELLA-II clinical trial. Patients received mid-position (midP) SBRT. The daily adapt-to-position workflow included localization, verification and intra-fraction tumor midP monitoring using 4D-MRI. Technical feasibility was established based on persistence of the treatment protocol, treatment time ≤1 h, no geographical miss and no unexpected acute toxicity grade >3. All 4D-MRIs were registered to the planning midP-CT and tumor midP and amplitude were calculated. Additionally, delivered target dose was accumulated incorporating the 4D-MRI intra-fraction tumor motion and evaluated with Monte-Carlo error simulations.
RESULTS: 20 patients with liver metastases were included and treated with 4D-MRI guided SBRT. Feasibility criteria were met in all-but-one patient. No grade ≥3 acute toxicity was observed. Group mean (M), systematic and random midP-drifts were 2.4 mm, 2.6 mm and 3.1 mm in CC-direction. 4D-MRI tumor CC-amplitudes were reduced compared to the simulation 4D-CT (M = -1.9 mm) and decreased during treatment (M = -1.4 mm). Dose accumulation showed adequate target-coverage on a population level.
CONCLUSION: We successfully demonstrated technical feasibility of 4D-MRI guided SBRT in a cohort of 20 patients with liver metastases. However, substantial midposition drifts occurred which stress the need for intra-fraction motion management strategies to further increase the precision of treatment delivery.