Quantification of Idelalisib in Human Plasma by Ultra-Performance Liquid Chromatography Coupled to Mass Spectrometry in Negative Ionization Mode.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Huynh Huu H, Roessle Clara, Sauvageon Hélène, Plé Alain, Madelaine Isabelle, Thieblemont Catherine, Mourah Samia, Goldwirt Lauriane
Аннотация
BACKGROUND: Idelalisib is the first orally active selective phosphatidylinositol 3-kinase delta inhibitor approved by Food and Drug Administration and European Medicines Agency in 2014 for the treatment of several types of blood cancer. Idelalisib is widely used as a monotherapy or in combination with rituximab, bendamustine, or ofatumumab with a significant efficacy. However, idelalisib has shown increased risk of infection and a higher frequency of serious adverse events. It may be useful to determine idelalisib concentration in human plasma to adjust dose and to manage adverse effects in clinical practice.
METHODS: After a single-step protein precipitation of plasma samples, the chromatographic separation was performed using an ultra-high performance liquid chromatography system coupled with mass tandem spectrometry in a negative ionization mode using isotope-labeled internal standard. This method was validated by studies of its linearity, accuracy, imprecision, limit of quantification, recovery, matrix effect, selectivity, and stability.
RESULTS: The quantification method was linear from 10 to 2500 ng/mL with a 5 ng/mL lower limit of quantification that encompasses the clinical range of drug concentration. The intraday and interday imprecisions were below 8.1% and 11.4%, respectively. The recoveries and matrix effect of idelalisib were 85.6% ± 1.2% and 95.7% ± 3.0%, respectively, which are consistent, precise, and reproducible (coefficient of variation % < 15%). Peak plasma concentration and trough plasma concentration ranges of idelalisib reached 1591-1937 ng/mL and 256.3-303.3 ng/mL, respectively, in 3 follicular lymphoma patients treated with idelalisib 150 mg twice a day.
CONCLUSIONS: A robust and sensitive liquid chromatography-tandem mass spectrometry method was developed and validated to quantify idelalisib concentration in human plasma. This method was effectively applied to 3 follicular lymphoma patients.
METHODS: After a single-step protein precipitation of plasma samples, the chromatographic separation was performed using an ultra-high performance liquid chromatography system coupled with mass tandem spectrometry in a negative ionization mode using isotope-labeled internal standard. This method was validated by studies of its linearity, accuracy, imprecision, limit of quantification, recovery, matrix effect, selectivity, and stability.
RESULTS: The quantification method was linear from 10 to 2500 ng/mL with a 5 ng/mL lower limit of quantification that encompasses the clinical range of drug concentration. The intraday and interday imprecisions were below 8.1% and 11.4%, respectively. The recoveries and matrix effect of idelalisib were 85.6% ± 1.2% and 95.7% ± 3.0%, respectively, which are consistent, precise, and reproducible (coefficient of variation % < 15%). Peak plasma concentration and trough plasma concentration ranges of idelalisib reached 1591-1937 ng/mL and 256.3-303.3 ng/mL, respectively, in 3 follicular lymphoma patients treated with idelalisib 150 mg twice a day.
CONCLUSIONS: A robust and sensitive liquid chromatography-tandem mass spectrometry method was developed and validated to quantify idelalisib concentration in human plasma. This method was effectively applied to 3 follicular lymphoma patients.