Optimal collimator rotation based on the outline of multiple brain targets in VMAT.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kim Jung-In, Ahn Beom Seok, Choi Chang Heon, Park Jong Min, Park So-Yeon

Аннотация

BACKGROUND: The aim of this study was to investigate the dosimetric quality in volumetric modulated arc therapy (VMAT) plans with optimal collimator angles that can represent the outline of multiple brain targets.

METHODS: Twenty patients with multiple target volumes in the brain cases were selected retrospectively. To better represent the outline of the multiple brain targets, four conformal arc plans were generated for each patient using one full arc with four collimator settings. The optimal collimator angles calculated from the integrated multi-leaf collimator (MLC) aperture that had the smallest aperture size for certain collimator settings of the conformal arc plan were selected. VMAT plans with the optimal collimator angles with angular sections of 40° and 60° (Colli-VMAT (40°), Colli-VMAT (60°)) were generated, followed by evaluation of field sizes, dose-volumetric parameters and total monitor units (MUs).

RESULTS: Patient-averaged values of field sizes for Colli-VMAT (40°) (111.5 cm2) were lowest and 1.3 times smaller than those for Std-VMAT (143.6 cm2). Colli-VMAT plans improved sparing of most normal organs but for brain stem and left parotid gland. For the total MUs, the averaged values obtained with the Colli-VMAT (40°) (390 ± 148 MU) were smaller than those obtained with the Std-VMAT (472 ± 235 MU).

CONCLUSIONS: The Colli-VMAT plans with smaller angular sections could be suitable in the clinic for multiple brain targets as well as for irregularly shaped targets. Determination of the optimal collimator rotation generally showed good normal tissue sparing and MU reduction for multiple brain targets.