Implementation of a fully implantable middle-ear hearing device chip.

Lee Jyung Hyun, Kim Dong Wook, Seong Ki Woong, Kim Myoung Nam, Cho Jin-Ho

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVE

Recently, with the increase in the population of hearing impaired people, various types of hearing aids have been rapidly developed. In particular, a fully implantable middle ear hearing device (F-IMEHD) is developed for people with sensorineural hearing loss. The F-IMEHD system comprises an implantable microphone, a transducer, and a signal processor. The signal processor should have a small size and consume less power for implantation in a human body.

METHODS

In this study, we designed and fabricated a signal-processing chip using the modified FFT algorithm. This algorithm was developed focusing on eliminating time delay and system complexity in the transform process. The designed signal-processing chip comprises a 4-channel WDRC, a fitting memory, a communication 1control part, and a pulse density modulator. Each channel is separated using a 64-point fast Fourier transform (FFT) method and the gain value is matched using the fitting table in the fitting memory.

RESULTS AND

CONCLUSION

The chip was designed by Verilog-HDL and the designed HDL codes were verified by Modelsim-PE 10.3 (Mentor graphics, USA). The chip was fabricated using a 0.18 μm CMOS process (SMIC, China). Experiments were performed on a cadaver to verify the performance of the fabricated chip.

Для цитирования: Lee Jyung Hyun, Kim Dong Wook, Seong Ki Woong, Kim Myoung Nam, Cho Jin-Ho. Implementation of a fully implantable middle-ear hearing device chip.. Technology and health care : official journal of the European Society for Engineering and Medicine. 2021. DOI: 10.3233/thc-218038.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.