Assessment of preferable site for temperature measurement using non-contact infra-red temperature among pediatric patients.

Kavthekar Saiprasad O, Chougule Ashok A, Kavthekar Swati S, Kurane Anil B, Verma Sachin

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Accurate temperature measurement with little or no discomfort that is safe, without risk of hospital-acquired infections or perforations, is the preferred choice of medical professionals in pediatric settings. The objective was to discover the preferable site for body temperature measurement using non-contact infra-red thermometer (NCIT) among pediatric patients.

METHODS

NCIT measurement at mid forehead (F), right temporal region (T), right side of neck-over-carotid artery area (N), jugular notch (J), sternum (S), umbilical region and sublingual region (U) were compared with digital axillary temperature (DAT) in a single attempt in 500 patients, aged between 2 to 5 years with fever. Data was analyzed using Pearson's correlation, paired t-test and Bland-Altman plot to assess the correlation and agreement between the DAT and NCIT sites.

RESULTS

The mean temperature of NCIT-T (38.42±0.64 °C) was more agreeable with DAT (38.42±0.63 °C) compared to other body sites. The minimum mean bias of -0.00480 °C was noted for NCIT-F with 95% CI of -0.164-0.15; however, NCIT-F revealed many outliers as compared to NCIT-J. A strong positive correlation existed between DAT and NCIT sites (r value: 0.99-0.98). However, significant difference was found between DAT and NCIT-N, NCIT-F and NCIT-U (P<0.0001).

CONCLUSIONS

NCIT-J is the most preferable choice for measuring body temperature and can be interchanged with DAT. It will help to deliver fast results with enhanced patient comfort due to its non-invasive nature.

Для цитирования: Kavthekar Saiprasad O, Chougule Ashok A, Kavthekar Swati S, Kurane Anil B, Verma Sachin. Assessment of preferable site for temperature measurement using non-contact infra-red temperature among pediatric patients.. Minerva pediatrics. 2020. DOI: 10.23736/s2724-5276.20.05755-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.