A 60-channel high-density flexible receive array for pediatric abdominal MRI.

Lee Wonje, Stickle Yunjeong, Follante Clyve, Grafendorfer Thomas, Yang Taeyoung, Robb Fraser, Zhang Fan, Pauly John, Scott Greig, Vasanawala Shreyas, Syed Ali

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Conventional MRI coils offer suboptimal parallel imaging performance for young children. Our goal was to enhance imaging acceleration by dedicated, flexible high-density coil design for pediatric patients at 3T.

METHODS

We design, construct, and evaluate a highly flexible small loop array. Key design notes include full-wave simulation and analysis of the dual-turn loop, miniature feedboard allocation at the loop center, and cable management. Phantom experiments and adult and pediatric volunteer case studies were conducted to evaluate the small loop array imaging performance compared to commercial reference coils.

RESULTS

Dual-turn loop configuration forms higher preamp decoupling impedance than the same size single-turn, supporting a flexible form factor that requires a wide range of critical overlap. Both phantom and in-vivo studies demonstrate superior parallel imaging performance or high spatial resolution imaging using the small loop, compared to commercial reference coils.

CONCLUSION

A dedicated high-density coil array with a minimum inter-component interference layout design allows a flexible form factor and higher imaging accelerations. Phantom and in-vivo volunteer case studies demonstrate promising results in improving efficiency for pediatric patients in routine clinical imaging procedures.

Для цитирования: Lee Wonje, Stickle Yunjeong, Follante Clyve, Grafendorfer Thomas, Yang Taeyoung, Robb Fraser, Zhang Fan, Pauly John, Scott Greig, Vasanawala Shreyas, Syed Ali. A 60-channel high-density flexible receive array for pediatric abdominal MRI.. Magnetic resonance in medicine. 2025. DOI: 10.1002/mrm.30456.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.