Adverse impact of multileaf collimator field shaping on lens dose in children with acute leukemia receiving cranial irradiation.

Kalapurakal J A, Sathiaseelan V, Bista T, Marymont M H

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: This study was designed to investigate the impact of multileaf collimator (MLC) on lens dose in children with leukemia undergoing cranial irradiation.

METHODS AND MATERIALS: This is a prospective study utilizing three common cranial irradiation techniques. Technique A uses a half-beam, nondivergent radiation field. Technique B has the anterior divergent field edge at the lateral bony canthus. Technique C is similar to B, but with a field collimator angle. Thermoluminescent dosimeter (TLD) lens dose measurements were obtained in children and phantom with all three techniques.

RESULTS

Seventeen children were studied. Lens dose measurements were obtained in 14 children with technique A using MLC and blocks. In 7 of 14 children, dose measurements were obtained with MLC only. One child was treated with technique B and 2 children were treated with C, with MLC +/- blocks. In all 3 techniques, with MLC alone, the lens dose increased by 64%, 119%, and 72%, respectively. Similar results were obtained in phantom.

CONCLUSION

This study demonstrates that independent of irradiation technique, additional custom blocking is required to maximally protect the lens with MLC shaped fields. This is due to the lack of conformity between MLC and the desired field edge at the lateral bony canthus.

Для цитирования: Kalapurakal J A, Sathiaseelan V, Bista T, Marymont M H. Adverse impact of multileaf collimator field shaping on lens dose in children with acute leukemia receiving cranial irradiation.. International journal of radiation oncology, biology, physics. 2000. DOI: 10.1016/s0360-3016(00)00723-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.