DOI 10.1213/01.ane.0000278734.34434.60
A two-dimensional model of anatomic relationships during laryngoscopy.
Аннотация
BACKGROUND
The view obtained during direct laryngoscopy is only seen by a single anesthesiologist. The inability of instructors to observe the view poses problems for teaching the technique. The anatomic interactions affecting laryngoscopy are largely internal, hampering efforts to understand why some patients are unexpectedly difficult to intubate. In response, we have constructed a full scale, adjustable, two-dimensional model showing the head and neck in the sagittal plane. In this article, we validate the mannequin and test how various conditions or changes in equipment affect the laryngoscopic view.
METHODS
Model parameters were compared with literature values. Glottic exposure was evaluated over a range of jaw lengths and interincisor gaps for Macintosh 3, Miller 2, and Macintosh 4 blades.
RESULTS
Thirty segmental airway distances and 10 angles were within 1 standard deviation from published values. Spine and jaw mobilities approximated normal range of motion. Glottic exposure decreased steeply for mouth openings below a threshold. A larger mouth opening was required to obtain a view when the mandible was short. None of the blades exposed the glottis when mouth opening was narrow, 2.4 cm. The Macintosh 4 blade was closest to success, within 7 mm of viewing the posterior cords.
CONCLUSIONS
The model reflects an average 16-yr-old male patient in size, proportion, and mobility. It can be used to explicate how anatomic relationships affect laryngoscopy. An objective assessment is necessary to determine the model's utility for teaching and as a tool for researching the mechanisms responsible for laryngoscopic difficulty.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.