DOI 10.1002/acm2.14407

Медицинская библиотека · Journal of applied clinical medical physics · 2024

The delivery assessment for small targets on Halcyon radiotherapy system - Measured and calculated dose comparison.

Lankinen Linda, Kulmala Antti, Lehtomäki Jouko, Harju Ari

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

With the ever-increasing requirements of accuracy and personalization of radiotherapy treatments, stereotactic radiotherapy (SRT) with volumetric modulated arc therapy (VMAT) on O-ring Halcyon radiotherapy system could potentially provide a fast, safe, and feasible treatment option.

PURPOSE: The purpose of this study was to assess the delivery of Halcyon VMAT plans for small targets.

METHODS

Well-defined VMAT-SRT plans were created on Halcyon radiotherapy system with the stacked and staggered dual-layer MLC design for the film measurement set-up and the target sizes and shapes designed to emulate the targets of the stereotactic treatments. The planar dose distributions were acquired with film measurements and compared to a current clinical reference dose calculation with AcurosXB (v18.0, Varian Medical Systems) and to Monte Carlo simulations. With the collapsed arc versions of the VMAT-SRT plans, the uncertainty in dose delivery due to the multileaf collimator (MLC) without the gantry rotation could be separated and analyzed.

RESULTS

The target size was mainly limited by the resolution originated from the design of the MLC leaves. The results of the collapsed arc versions of the plans show good consistency among measured, calculated, and simulated dose distributions. With the full VMAT plans, the agreement between calculated and simulated dose distributions was consistent with the collapsed arc versions. The measured dose distribution agreed with the calculated and simulated dose distributions within the target regions, but considerable local differences were observed in the margins of the target. The largest differences located in the steep gradient regions presumably originating from the deviation of the isocenter.

CONCLUSIONS

The potential of the Halcyon radiotherapy system for VMAT-SRT delivery was evaluated and the study revealed valuable insights on the machine characteristics with the delivery.

Для цитирования: Lankinen Linda, Kulmala Antti, Lehtomäki Jouko, Harju Ari. The delivery assessment for small targets on Halcyon radiotherapy system - Measured and calculated dose comparison.. Journal of applied clinical medical physics. 2024. DOI: 10.1002/acm2.14407.