Metal and motion artifacts by cone beam computed tomography (CBCT) in dental and maxillofacial study.

Nardi Cosimo, Borri Claudia, Regini Francesco, Calistri Linda, Castellani Alessandro, Lorini Chiara, Colagrande Stefano

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study aimed at evaluating incidence/degree of metal/motion artifacts and CT-dose-index in oral/maxillofacial examinations using Cone-Beam-CT.

METHODS

Interferences caused by metal and motion artifacts were evaluated in 500 patients aged from 6 to 81 years, in dental arches, maxillofacial and splanchocranium Cone-Beam-CT exams. The interferences was divided into four progressive degrees (G0-G3) related to the possibility to answer the clinical query. The parameters considered were field-of-view, scan time, patient's age, and anatomical area. Furthermore volumetric CT-dose-index was measured.

RESULTS

In the presence of metal artifacts the clinical query was always answered (G3 = 0). No artifacts (G0) were found in all cases when metal was beyond 5 cm from interest site and in 18.4% when metal was inside this distance. Beam hardening and photon starvation due to implants, restoration and orthodontic therapies achieved 56.2% G1 and 25.4% G2. Motion artifacts were more frequent in under ten (31.5%) and over sixty (82.2%), and in mandible analysis (inferior arch 59.5%, both arches 47.3%). Moreover, their incidence and intensity were influenced by scan time (49.1% at 36 s) but not by field-of-view. Mean volumetric CT-dose-index of all patients was mGy 9.11 (mGy 3.62, 5.78, 8.89, and 13.07 at 18, 24, 26, and 36 s, respectively).

CONCLUSIONS

In our series Cone-Beam-CT diagnostic evaluation was never inhibited by metal artifacts and only in 1.9% of the cases by motion artifacts, always with a very low CT-dose-index.

Для цитирования: Nardi Cosimo, Borri Claudia, Regini Francesco, Calistri Linda, Castellani Alessandro, Lorini Chiara, Colagrande Stefano. Metal and motion artifacts by cone beam computed tomography (CBCT) in dental and maxillofacial study.. La Radiologia medica. 2015. DOI: 10.1007/s11547-015-0496-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.