Real-time bowel perfusion monitoring with FUJIFILM ELUXEO® VISION endoscopic imaging system in colorectal surgery.

Martucci Jordan, Williams Brian, Swinford Aubrey, Gupta Abhinav, Peters Alexis, Lim Anthony, Broccard Sacha, Cologne Kyle G, Shin Joongho, Lee Sang W

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Anastomotic leak is a serious complication of bowel anastomosis, with multifactorial etiology. Thorough intraoperative evaluation of the anastomosis using white light endoscopy and leak tests is critical. Indocyanine green fluorescence angiography (ICG-FA) has emerged as a tool to assess bowel perfusion, but has limitations, including patient allergies and procedural complexity. The FUJIFILM ELUXEO® VISION Endoscopic Imaging System uses oxygen saturation imaging (OSI) to measure tissue oxygen saturation (StO2) in real time, without requiring dye injection. This study investigates a novel quantitative approach for assessing bowel perfusion endoscopically by evaluating StO2 at colon anastomoses.

METHODS

The FUJIFILM VISION was used to obtain intraoperative mucosal StO2 measurements in 12 patients' post-colon resection with anastomosis. Measurements were taken at four locations: proximal base, proximal staple line, distal staple line, and distal base. Two-tailed paired t tests were used to compare StO2 measurements at distant healthy mucosa (base) to the staple line, both proximal and distal to the anastomosis. Patients were followed longitudinally via chart review to monitor outcomes.

RESULTS

The average StO2 differences were + 12.58% (95% CI 1.96 - 23.19%) between the proximal base and proximal staple line and + 15.34% (95% CI 6.76 - 23.9%) between the distal base and distal staple line. Significant StO2 differences were observed between baseline normal mucosa and staple line adjacent mucosa (proximal: p = 0.025; distal: p = 0.003). No patients had intraoperative complications or developed anastomotic leaks during the study.

CONCLUSIONS

This study shows that the ELUXEO® VISION Endoscopic Imaging System is a potentially useful tool for providing quantitative data on tissue perfusion at and near anastomotic sites, complementing traditional white light endoscopy and leak tests. Larger prospective studies with long-term follow-up are needed to confirm the relationship between StO2 measurements and clinical outcomes such as anastomotic leaks and stricture.

Для цитирования: Martucci Jordan, Williams Brian, Swinford Aubrey, Gupta Abhinav, Peters Alexis, Lim Anthony, Broccard Sacha, Cologne Kyle G, Shin Joongho, Lee Sang W. Real-time bowel perfusion monitoring with FUJIFILM ELUXEO® VISION endoscopic imaging system in colorectal surgery.. Surgical endoscopy. 2025. DOI: 10.1007/s00464-025-12526-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.