MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcms.2013.10.001

Медицинская библиотека · Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery · 2013

Inaccuracies in additive manufactured medical skull models caused by the DICOM to STL conversion process.

Huotilainen Eero, Jaanimets Risto, Valášek Jiří, Marcián Petr, Salmi Mika, Tuomi Jukka, Mäkitie Antti, Wolff Jan

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

The process of fabricating physical medical skull models requires many steps, each of which is a potential source of geometric error. The aim of this study was to demonstrate inaccuracies and differences caused by DICOM to STL conversion in additively manufactured medical skull models.

MATERIAL AND

METHODS

Three different institutes were requested to perform an automatic reconstruction from an identical DICOM data set of a patients undergoing tumour surgery into an STL file format using their software of preference. The acquired digitized STL data sets were assessed and compared and subsequently used to fabricate physical medical skull models. The three fabricated skull models were then scanned, and differences in the model geometries were assessed using established CAD inspection software methods.

RESULTS

A large variation was noted in size and anatomical geometries of the three physical skull models fabricated from an identical (or "a single") DICOM data set.

CONCLUSIONS

A medical skull model of the same individual can vary markedly depending on the DICOM to STL conversion software and the technical parameters used. Clinicians should be aware of this inaccuracy in certain applications.

Для цитирования: Huotilainen Eero, Jaanimets Risto, Valášek Jiří, Marcián Petr, Salmi Mika, Tuomi Jukka, Mäkitie Antti, Wolff Jan. Inaccuracies in additive manufactured medical skull models caused by the DICOM to STL conversion process.. Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery. 2013. DOI: 10.1016/j.jcms.2013.10.001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление