DOI 10.1002/jcla.22234

Медицинская библиотека · Journal of clinical laboratory analysis · 2017

Comparison of three rapid influenza diagnostic tests with digital readout systems and one conventional rapid influenza diagnostic test.

Ryu Sook Won, Suh In Bum, Ryu Se-Min, Shin Kyu Sung, Kim Hyon-Suk, Kim Juwon, Uh Young, Yoon Kap Jun, Lee Jong-Han

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Rapid influenza diagnostic tests (RIDTs) show variable sensitivities in clinical settings. We aimed to compare three digital RIDTs and one conventional RIDT.

METHODS

We assessed 218 nasopharyngeal swabs from patients between neonates and 90 years old in 2016. Three digital RIDTs were BUDDI, Sofia Influenza A+B Fluorescence Immunoassay, Veritor System Flu A+B assay. One conventional test was the SD Bioline Influenza Ag A/B/A(H1N1/2009). All test results were compared with those from the Anyplex Flu A/B Typing Real-time Detection real-time PCR. The four RIDTs were tested with diluted solutions from the National Institute for Biological Standards and Control (NIBSC) to compare lower detection limit. Cross-reactivity of four RIDTs within other respiratory viruses was identified.

RESULTS

For influenza A, BUDDI, Sofia, Veritor, and Bioline showed 87.7%, 94.5%, 87.7%, and 72.6% sensitivity, and 100%, 97.7%, 96.5%, and 100% specificity. For influenza B, BUDDI, Sofia, Veritor, and Bioline showed 81.7%, 91.7%, 81.7%, and 78.3% sensitivity, and 100%, 95.3%, 100%, and 100% specificity, respectively. Each RIDT could detect diluted NIBSC solution, according to the level of dilution and specific influenza subtypes. Cross-reactivity of four RIDTs with other respiratory viruses was not noted.

CONCLUSIONS

Sofia showed the highest sensitivity for influenza A and B detection. BUDDI and Veritor showed higher detection sensitivity than a conventional RIDT for influenza A detection, but similar results for influenza B detection. Further study is needed to compare the test performance of RIDTs according to specific, prevalent influenza subtypes.

Для цитирования: Ryu Sook Won, Suh In Bum, Ryu Se-Min, Shin Kyu Sung, Kim Hyon-Suk, Kim Juwon, Uh Young, Yoon Kap Jun, Lee Jong-Han. Comparison of three rapid influenza diagnostic tests with digital readout systems and one conventional rapid influenza diagnostic test.. Journal of clinical laboratory analysis. 2017. DOI: 10.1002/jcla.22234.

Предметные рубрики MeSH