DOI 10.1016/j.mri.2012.09.008

Медицинская библиотека · Magnetic resonance imaging · 2012

MR elastography of the head and neck: driver design and initial results.

Yeung David K W, Bhatia Kunwar S, Lee Yolanda Y P, King Ann D, Garteiser Philippe, Sinkus Ralph, Ahuja Anil T

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: The purpose was to describe the design and fabrication of a driver suitable for magnetic resonance elastography (MRE) of the head and neck and to assess its performance in evaluating human parotid gland, lymph nodes and thyroid at 3.0 T.

MATERIALS AND

METHODS

A head and neck driver was fabricated using a commercial transducer, headrest mould and piston extension. Driver performance was tested using a motion-sensitized spin-echo MRE pulse sequence. Six healthy volunteers and three patients (two metastatic nodes and one papillary carcinoma) were evaluated using MRE. Viscoelastic maps were computed to obtain storage modulus (G') and loss modulus (G") of the normal parotid and thyroid, metastatic node and thyroid cancer. Reproducibility was assessed by coefficient of variation.

RESULTS

All subjects completed MRE examination without discomfort. Initial G' and G" values were as follows: normal parotid gland, 1.12 kPa and 0.48 kPa; thyroid, 0.58 kPa and 0.42 kPa; metastatic node, 0.66 kPa and 0.58 kPa; and thyroid cancer, 0.17 kPa and 0.28 kPa. Based on parotid data, the coefficient of variation for G' and G" was 4.7% and 9.8%.

CONCLUSION

A new MRE driver for head and neck was successfully implemented, and our initial results suggested the device was suitable for the mechanical assessment of tissues in the head and neck.

Для цитирования: Yeung David K W, Bhatia Kunwar S, Lee Yolanda Y P, King Ann D, Garteiser Philippe, Sinkus Ralph, Ahuja Anil T. MR elastography of the head and neck: driver design and initial results.. Magnetic resonance imaging. 2012. DOI: 10.1016/j.mri.2012.09.008.