DOI 10.1097/mao.0000000000004533

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2025

Clinical Evidence on the Influence of Implant Position onto Maximum Output with the Bonebridge Bone Conduction Implant.

Schurzig Daniel, Iseke Rebecca, Maier Hannes, Prenzler Nils Kristian, Lenarz Thomas, Ghoncheh Mohammad

СтатьяПеревод в очереди

Аннотация

HYPOTHESIS: In bone conduction implantation, the position of the implant influences the audiological benefit of the patient.

BACKGROUND

One way of treating hearing loss is the implantation of bone conduction implants (BCIs), which effectively transmit vibrations through the skull bone to the cochlea given that the implant transducer is securely fixated. Laboratory research on the efficacy of bone conduction sound transmission found that a closer proximity of the transducer to the ipsilateral cochlea yields significantly higher cochlear promontory vibrations and hence, higher stimulation efficacy. Up to now, this finding has not been reproduced using clinical data such as the functional or effective gain.

METHODS

The present, retrospective study was conducted on a cohort of 28 BCI patients to correlate the implantation site of the BC transducer, derived from clinical postoperative imaging and defined in a standardized coordinate system, with maximum output values that are exclusively based on a novel calculation method only employing clinical audiological data.

RESULTS

It could be shown that the efficacy of BCI stimulation is in fact correlated with the transducer distance to the cochlea, and that this correlation is frequency dependent. Furthermore, the longitudinal distance of the transducer and the ipsilateral external auditory canal is negatively correlated with the maximal output while the sagittal distance is not.

CONCLUSION

The present study is hence the first one to clinically demonstrate the significance of BCI placement for maximizing patient benefit, which should be considered during the preoperative planning of bone conduction implantation.

Для цитирования: Schurzig Daniel, Iseke Rebecca, Maier Hannes, Prenzler Nils Kristian, Lenarz Thomas, Ghoncheh Mohammad. Clinical Evidence on the Influence of Implant Position onto Maximum Output with the Bonebridge Bone Conduction Implant.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2025. DOI: 10.1097/mao.0000000000004533.