Intracuff pressure during one-lung ventilation in infants and children.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Shafy Shabana Z, Hakim Mohammed, Kamata Mineto, Tumin Dmitry, Krishna Senthil G, Naguib Aymen, Tobias Joseph D
Аннотация
OBJECTIVE: We prospectively evaluated intracuff pressure (IP) during one-lung ventilation (OLV) to characterize potential risk associated with overinflation of the cuff used for OLV.
DESIGN: Prospective observational study over a 2-year period, in infants and children undergoing thoracic surgery. The IPs of the tracheal and bronchial balloon were measured using a manometer and compared to a previously recommended threshold of 30 cmH2O. Data were compared by the device type used to achieve OLV.
SETTING: Freestanding tertiary-care pediatric hospital.
PARTICIPANTS: Patients ≤18 years of age undergoing thoracic procedures requiring OLV.
INTERVENTIONS: Measurement of IP.
MEASUREMENTS AND MAIN RESULTS: Thirty patients were enrolled (age 5 months-18 years) with a median weight of 28 kg. Median tracheal and bronchial IPs were 32 cmH2O (range: 11, 90) and 44 cmH2O (range: 10, 100), respectively. The tracheal and bronchial IPs exceeded 30 cmH2O in 13 of 20 patients (65%) and 21 of 30 patients (70%), respectively.
CONCLUSIONS: IP was high and in excess of recommended levels in most children undergoing OLV. Continuous monitoring of IP may be indicated during OLV to address the risks involved and ensure the prevention of complications related to high IP.
TYPE OF STUDY: Prospective comparative study.
LEVEL OF EVIDENCE: Level II.
DESIGN: Prospective observational study over a 2-year period, in infants and children undergoing thoracic surgery. The IPs of the tracheal and bronchial balloon were measured using a manometer and compared to a previously recommended threshold of 30 cmH2O. Data were compared by the device type used to achieve OLV.
SETTING: Freestanding tertiary-care pediatric hospital.
PARTICIPANTS: Patients ≤18 years of age undergoing thoracic procedures requiring OLV.
INTERVENTIONS: Measurement of IP.
MEASUREMENTS AND MAIN RESULTS: Thirty patients were enrolled (age 5 months-18 years) with a median weight of 28 kg. Median tracheal and bronchial IPs were 32 cmH2O (range: 11, 90) and 44 cmH2O (range: 10, 100), respectively. The tracheal and bronchial IPs exceeded 30 cmH2O in 13 of 20 patients (65%) and 21 of 30 patients (70%), respectively.
CONCLUSIONS: IP was high and in excess of recommended levels in most children undergoing OLV. Continuous monitoring of IP may be indicated during OLV to address the risks involved and ensure the prevention of complications related to high IP.
TYPE OF STUDY: Prospective comparative study.
LEVEL OF EVIDENCE: Level II.