Decrease the flow setting to improve trans-nasal pulmonary aerosol delivery via "high-flow nasal cannula" to infants and toddlers.

Li Jie, Gong Lingyue, Ari Arzu, Fink James B

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Trans-nasal pulmonary aerosol delivery for infants and toddlers has recently gained popularity, however, the reported lung deposition is low. We aimed to investigate the influential factors to improve the delivery.

METHODS

Anatomic airway manikins simulating infant (5 kg) and toddler (15 kg) with collecting filter connected the trachea and breath simulator, were set to represent quiet and distressed breathing. Nasal cannula flow was set at 0.125, 0.25, 0.5, 1, and 2 L/kg/min. A mesh nebulizer (Aerogen) was placed at the inlet of humidifier (Fisher & Paykel) and proximal to patient. Albuterol (5 mg in 1 mL) was nebulized for each condition (n = 3). Drug was eluted from the filter and assayed with UV spectrophotometry (276 nm).

RESULTS

Inhaled dose was higher with nebulizer placed at the inlet of humidifier than proximal to patient in all settings, except the infant model at low gas flow settings (0.125 and 0.25 L/kg/min). When nebulizer was placed at the inlet of humidifier, inhaled dose was higher when gas flow was below patient's inspiratory flow than when gas flow exceeded patient's inspiratory flow (8.77 ± 3.84 vs 2.16 ± 1.29%, P < 0.001); aerosol deposition increased as gas flow decreased, with greatest deposition at gas flow of 0.25 L/kg/min (11.29 ± 2.15%). A multiple linear regression identified gas flow as the primary predictor of aerosol delivery.

CONCLUSIONS

Trans-nasal pulmonary aerosol delivery was significantly improved when gas flow was below patient's inspiratory flow, aerosol deposition increased with decreased nasal cannula flow, with greatest deposition at 0.25 L/kg/min.

Для цитирования: Li Jie, Gong Lingyue, Ari Arzu, Fink James B. Decrease the flow setting to improve trans-nasal pulmonary aerosol delivery via "high-flow nasal cannula" to infants and toddlers.. Pediatric pulmonology. 2019. DOI: 10.1002/ppul.24274.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.