Validation of a new transit time ultrasound flowmeter for measuring blood flow in colonic mesenteric arteries.

Mortensen F V, Rasmussen J S, Viborg O, Laurberg S, Pedersen E M

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To validate a transit time ultrasound flowmeter (CardioMed CM 4000) for measuring blood flow in isolated colonic mesenteric arteries.

DESIGN

Experimental and clinical study.

SETTING

Teaching hospital, Denmark. ANIMALS AND SUBJECTS: One female pig, and 6 patients being operated on for carcinoma of the sigmoid colon and rectum.

INTERVENTIONS: Volume blood flow measured by Cardiomed CM 4000 and pump withdrawal flow recordings.

MAIN OUTCOME MEASURES: Correlation between the two methods.

RESULTS

There was good agreement between transit time flow recordings and pump withdrawal flow recordings (correlation coefficient of 1.0). Of the differences between the two methods, 95% were between -0.16 ml min(-1) and 1.29 ml min(-1), mean 0.57 ml min(-1), or (in percentages) 105, 95% lying between 97-115. There was also good reproducibility in transit time flow recordings, the mean difference between repeated measurements being 0.06 ml min(-1), 95% lying between -0.66 ml min(-1) and 0.78 ml min(-1).

CONCLUSION

Ultrasound transit time flow recordings gave precise measurements of blood flow in isolated colonic mesenteric arteries.

Для цитирования: Mortensen F V, Rasmussen J S, Viborg O, Laurberg S, Pedersen E M. Validation of a new transit time ultrasound flowmeter for measuring blood flow in colonic mesenteric arteries.. The European journal of surgery = Acta chirurgica. 1998. DOI: 10.1080/110241598750005705.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.