Intensity modulated high dose rate ocular brachytherapy using Se-75.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Dupere Justine M, Munro John J, Medich David C

Аннотация

PURPOSE: We propose an alternative to LDR brachytherapy for the treatment of ocular melanomas by coupling intensity modulation, through the use of a gold shielded ring applicator, with a middle energy HDR brachytherapy source, Se-75. In this study, we computationally test this proposed design using MCNP6.

METHODS AND MATERIALS: An array of discrete Se-75 sources is formed into a ring configuration within a gold shielded applicator, which collimates the beam to a conical shape. Varying this angle of collimation allows for the prescription dose to be delivered to the apex of various sized targets. Simulations in MCNP6 were performed to calculate the dosimetric output of the Se-75 ring source for various sized applicators, collimators, and target sizes.

RESULTS: The prescription dose was delivered to a range of target apex depths 3.5-8 mm in the eye covering targets 10-15 mm in diameter by using various sized applicators and collimators. For a 16 mm applicator with a collimator opening that delivers the prescription dose to a depth of 5 mm in the eye, the maximum percent dose rate to critical structures was 30.5% to the cornea, 35.7% to the posterior lens, 33.3% to the iris, 20.1% to the optic nerve, 278.0% to the sclera, and 267.3% to the tumor.

CONCLUSIONS: When using Se-75 in combination with the proposed gold shielded ring applicator, dose distributions are appropriate for ocular brachytherapy. The use of a collimator allows for the dose to more easily conform to the tumor volume. This method also reduces treatment time and cost, and it eliminates hand dose to the surgeon through the use of a remote afterloader device.