DOI 10.1016/j.radonc.2023.109963

Медицинская библиотека · Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2023

Paediatric CBCT protocols for image-guided radiotherapy; outcome of a survey across SIOP Europe affiliated countries and literature review.

Østergaard Daniella Elisabet, Bryce-Atkinson Abigail, Skaarup Mikkel, Smulders Bob, Davies Lucy Siew Chen, Whitfield Gillian, Janssens Geert O, Hjalgrim Lisa Lyngsie, Richter Ivan Vogelius, van Herk Marcel, Aznar Marianne, Vestmø Maraldo Maja

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Implementation of daily cone-beam CT (CBCT) into clinical practice in paediatric image-guided radiotherapy (IGRT) lags behind compared to adults. Surveys report wide variation in practice for paediatric IGRT and technical information remains unreported. In this study we report on technical settings from applied paediatric CBCT protocols and review the literature for paediatric CBCT protocols.

METHODS

From September to October 2022, a survey was conducted among 246 SIOPE-affiliated centres across 35 countries. The survey consisted of 3 parts: 1) baseline information; technical CBCT exposure settings and patient set-up procedure for 2) brain/head, and 3) abdomen. Descriptive statistics was used to summarise current practice. The literature was reviewed systematically with two reviewers obtaining consensus

RESULTS

The literature search revealed 22 papers concerning paediatric CBCT protocols. Seven papers focused on dose-optimisation. Responses from 50/246 centres in 25/35 countries were collected: 44/50 treated with photons and 10/50 with protons. In total, 48 brain/head and 53 abdominal protocols were reported. 42/50 centres used kV-CBCT for brain/head and 35/50 for abdomen; daily CBCT was used for brain/head = 28/48 (58%) and abdomen = 33/53 62%. Greater consistency was seen in brain/head protocols (dose range 0.32 - 67.7 mGy) compared to abdominal (dose range 0.27 - 119.7 mGy).

CONCLUSION

Although daily CBCT is now widely used in paediatric IGRT, our survey demonstrates a wide range of technical settings, suggesting an unmet need to optimise paediatric IGRT protocols. This is in accordance with the literature. However, there are only few paediatric optimisation studies suggesting that dose reduction is possible while maintaining image quality.

Для цитирования: Østergaard Daniella Elisabet, Bryce-Atkinson Abigail, Skaarup Mikkel, Smulders Bob, Davies Lucy Siew Chen, Whitfield Gillian, Janssens Geert O, Hjalgrim Lisa Lyngsie, Richter Ivan Vogelius, van Herk Marcel, Aznar Marianne, Vestmø Maraldo Maja. Paediatric CBCT protocols for image-guided radiotherapy; outcome of a survey across SIOP Europe affiliated countries and literature review.. Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 2023. DOI: 10.1016/j.radonc.2023.109963.

Предметные рубрики MeSH