Comparative evaluation of the Diagenode multiplex PCR assay on the BD max system versus a routine in-house assay for detection of Bordetella pertussis.
Сравнительное исследованиеОценочное исследованиеСтатьяПеревод пока не подготовлен
Аннотация
This study looked at 128 nasopharyngeal aspirates (NPA) and 162 throat swabs (TS) tested with the Diagenode multiplex assay on the BD Max system versus our in-house Bordetella pertussis PCR. Sensitivity and specificity were 97.3% and 100% for NPA and 88.3% and 98% for TS, respectively. Of positive NPA, 42.1% were coinfected with respiratory viruses.
Для цитирования: Kenicer Juliet, Hardie Alison, Hamilton Fiona, Gadsby Naomi, Templeton Kate. Comparative evaluation of the Diagenode multiplex PCR assay on the BD max system versus a routine in-house assay for detection of Bordetella pertussis.. Journal of clinical microbiology. 2014. DOI: 10.1128/jcm.00212-14.