Performance evaluation of manual and automated (MagNA pure) nucleic acid isolation in HPV detection and genotyping using Roche Linear Array HPV Test.

Chranioti Aikaterini, Aga Evangelia, Margari Niki, Kottaridi Christine, Pappas Asimakis, Panayiotides Ioannis, Karakitsos Petros

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

Nucleic acids of human papillomavirus (HPV) isolated by manual extraction method (AmpliLute) and automated MagNA pure system were compared and evaluated with cytohistological findings in 253 women. The concordance level between AmpliLute and MagNA was very good 93.3% (κ = 0.864, P < .0001). Overall HPVpositivity detected by AmpliLute was 57.3% (30.4% as single and 27% as multiple infections) in contrast to MagNA 54.5% (32% and 23%, resp.). Discrepant results observed in 25 cases: 11 MagNA(-)/AmpliLute(+), 10 of which had positive histology; 5 MagNA(+)/AmpliLute(-) with negative histology; 8 MagNA(+)/AmpliLute(+): in 7 of which AmpliLute detected extra HPV genotypes and 1 MagNA(invalid)/AmpliLute(+) with positive histology. Both methods performed well when compared against cytological (area under curve (AUC) of AmpliLute 0.712 versus 0.672 of MagNA) and histological diagnoses (AUC of AmpliLute 0.935 versus 0.877 of MagNA), with AmpliLute showing a slightly predominance over MagNA. However, higher sensitivities, specificities, and positive/negative predictive values were obtained by AmpliLute.

Для цитирования: Chranioti Aikaterini, Aga Evangelia, Margari Niki, Kottaridi Christine, Pappas Asimakis, Panayiotides Ioannis, Karakitsos Petros. Performance evaluation of manual and automated (MagNA pure) nucleic acid isolation in HPV detection and genotyping using Roche Linear Array HPV Test.. Infectious diseases in obstetrics and gynecology. 2011. DOI: 10.1155/2011/931281.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.