In vivo diode dosimetry vs. computerized tomography and digitally reconstructed radiographs for critical organ dose calculation in high-dose-rate brachytherapy of cervical cancer.

Hassouna Ashraf H, Bahadur Yasir A, Constantinescu Camelia, El Sayed Mohamed E, Naseem Hussain, Naga Adly F

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate the correlation between the dose predicted by the treatment planning system using digitally reconstructed radiographs or three-dimensional (3D)-reconstructed CT images and the dose measured by semiconductor detectors, under clinical conditions of high-dose-rate brachytherapy of the cervix uteri.

PATIENTS AND

METHODS

Thirty-two intracavitary brachytherapy applications were performed for 12 patients with cancer of the cervix uteri. The prescribed dose to Point A was 7 Gy. Dose was calculated for both International Commissioning on Radiation Units and Measurements (ICRU) bladder and rectal points based on digitally reconstructed radiographs and for 3D CT images-based volumetric calculation of the bladder and rectum. In vivo diode dosimetry was performed for the bladder and rectum.

RESULTS

The ICRU reference point and the volumes of 1, 2, and 5cm(3) received 3.6±0.9, 5.6±2.0, 5.1±1.7, 4.3±1.4 and 5.0±1.2, 5.3±1.3, 4.9±1.1, and 4.2±0.9 Gy for the bladder and rectum, respectively. The ratio of the 1cm(3) and the ICRU reference point dose to the diode dose was 1.8±0.7 and 1.2±0.5 for the bladder and 1.9±0.6 and 1.7±0.5 for the rectum, respectively.

CONCLUSIONS

3D image-based dose calculation is the most accurate and reliable method to evaluate the dose given to critical organs. In vivo diode dosimetry is an important method of quality assurance, but clinical decisions should be made based on 3D-reconstructed CT image calculations.

Для цитирования: Hassouna Ashraf H, Bahadur Yasir A, Constantinescu Camelia, El Sayed Mohamed E, Naseem Hussain, Naga Adly F. In vivo diode dosimetry vs. computerized tomography and digitally reconstructed radiographs for critical organ dose calculation in high-dose-rate brachytherapy of cervical cancer.. Brachytherapy. 2011. DOI: 10.1016/j.brachy.2011.03.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.