Patient positioning during pediatric cardiothoracic computed tomography using a high-resilience pad system and pre-scan measurement of chest thickness.

Higuchi Satoshi, Nishii Tatsuya, Hirota Atsushi, Harumoto Shota, Horinouchi Hiroki, Tateishi Emi, Ohta Yasutoshi, Kiso Keisuke, Kurosaki Kenichi, Fukuda Tetsuya

СтатьяПеревод пока не подготовлен

Аннотация

Patient positioning at the isocenter of the CT gantry is important for optimizing image quality and radiation dose, but accurate positioning is challenging in pediatric patients. We evaluated whether the high-resilience pad and pre-scan measurement of chest thickness allow accurate positioning in pediatric patients with congenital heart disease. Sixty-seven patients aged 7 years or younger who underwent cardiothoracic CT were enrolled. The ideal table height, defined as the position at which the scanner's and patient's isocenters coincided, was determined by radiographers either manually (manual group) or based on the pad's and chest's thickness (calculated group). The distance between the two isocenters and image quality were evaluated. The calculated group demonstrated smaller isocenter distance and standard deviation (distance: 0.2 ± 5.8 mm vs. - 8.3 ± 11.6 mm, p < 0.01; absolute value: 4.1 [1.9-8.0] mm vs. 12.3 [5.1-16.3] mm, p < 0.01), and higher signal-to-noise ratio (SNR) and dose-normalized SNR (SNRD) in the descending aorta than the manual group (SNR: 39.8 [31.0-53.7] vs. 31.9 [28.9-36.6], p = 0.048, SNRD: 39.8 [31.0-53.7] vs. 31.9 [28.9-36.6], p = 0.04). The system allowed for more accurate positioning in pediatric cardiothoracic CT, yielding higher image quality.

Для цитирования: Higuchi Satoshi, Nishii Tatsuya, Hirota Atsushi, Harumoto Shota, Horinouchi Hiroki, Tateishi Emi, Ohta Yasutoshi, Kiso Keisuke, Kurosaki Kenichi, Fukuda Tetsuya. Patient positioning during pediatric cardiothoracic computed tomography using a high-resilience pad system and pre-scan measurement of chest thickness.. Scientific reports. 2022. DOI: 10.1038/s41598-022-21018-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.