MedElevator Войти в исследования
← Поиск публикаций

DOI 10.4103/0973-1482.119316

Медицинская библиотека · Journal of cancer research and therapeutics · 2013

Dosimetric comparison between three dimensional treatment planning system, Monte Carlo simulation and gel dosimetry in nasopharynx phantom for high dose rate brachytherapy.

Fazli Zeynab, Sadeghi Mahdi, Zahmatkesh M H, Mahdavi Seied Rabei, Tenreiro Claudio

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: For the treatment of nasopharnx carcinoma (NPC) using brachytherapy methods and high-energy photon sources are common techniques. In the common three dimensional (3D) treatments planning, all of the computed tomography images are assumed homogeneous. This study presents the results of Monte Carlo calculations for non-homogeneous nasopharynx phantom, MAGICA normoxic gel dosimetry and 3D treatment planning system (TPS).

MATERIALS AND

METHODS

The head phantom was designed with Plexiglas cylinder, head bone, and nasopharynx brachytherapy silicon applicator. For the simulations, version 5 of the Monte Carlo N-particle transport code (MCNP5) was used. 3D treatment planning was performed in Flexiplan software. A normoxic radiosensitive polymer gel was fabricated under normal atmospheric conditions and poured into test tubes (for calibration curve) and the head phantom. In addition, the head phantom was irradiated with Flexitron afterloader brachytherapy machine with (192)Ir source. To obtain calibration curves, 11 dosimeters were irradiated with dose range of 0-2000 cGy. Evaluations of dosimeters were performed on 1.5T scanner.

RESULTS

Two-dimensional iso-dose in coronal plan at distances of z = +0.3, -0.3 cm was calculated. There was a good accordance between 3D TPS and MCNP5 simulation and differences in various distances were between 2.4% and 6.1%. There was a predictable accordance between MAGICA gel dosimetry and MCNP5 simulation and differences in various distances were between 5.7% and 7.4%. Moreover, there was an acceptable accordance between MAGICA gel dosimetry and MCNP5 data and differences in various distances were between 5.2% and 9.4%.

CONCLUSION

The sources of differences in this comparison are divided to calculations variation and practical errors that was added in experimental dosimetry. The result of quality assurance of nasopharynx high dose rate brachytherapy is consistent with international standards.

Для цитирования: Fazli Zeynab, Sadeghi Mahdi, Zahmatkesh M H, Mahdavi Seied Rabei, Tenreiro Claudio. Dosimetric comparison between three dimensional treatment planning system, Monte Carlo simulation and gel dosimetry in nasopharynx phantom for high dose rate brachytherapy.. Journal of cancer research and therapeutics. 2013. DOI: 10.4103/0973-1482.119316.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.