MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1753-0407.2012.00192.x

Медицинская библиотека · Journal of diabetes · 2012

Safety and accuracy of a new long-term subconjunctival glucose sensor.

Hasslacher Christoph, Auffarth Gerd, Platten Isabell, Rabsilber Tanja, Smith Beate, Kulozik Felix, Knuth Monika, Nikolaus Katharina, Müller Achim

Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

A new biosensor has been developed by EyeSense (Großostheim, Germany) that is placed into the conjunctiva of one eye to measure the glucose concentration of the surrounding tissue in a non-invasive manner. In the present study we investigated the correlation between glucose concentrations measured by the EyeSense implant and those determined by finger prick testing, as well as the tolerability and safety of the implant over a 16-week period.

METHODS

The study was performed in 28 diabetic patients. The biosensor was inserted under local anesthesia and sterile conditions. Correlations between capillary glucose measured by laboratory methods and interstitial glucose determined by the biosensor were investigated by inducing increases and decreases in glucose values between 60 and 300 mg/dL.

RESULTS

Most patients experienced a mild subconjunctival hemorrhage postoperatively. Except for the minor sensation of the presence of foreign body, the implants were well tolerated. Three patients lost the ocular mini insert spontaneously, whereas there was a function failure of the insert in four patients. Error grid analysis showed that the percentage of data pairs in the acceptable ranges (zone A and B) was very high (>96%). However, there was a shift from zone A to zone B during observation. This was due primarily to an increase in the lag time between capillary and interstitial measured glucose.

CONCLUSION

he present study demonstrates good tolerability and measurement performance of the biosensor. The reasons for an increase in the lag time are still unknown; local reactions may be involved.

Для цитирования: Hasslacher Christoph, Auffarth Gerd, Platten Isabell, Rabsilber Tanja, Smith Beate, Kulozik Felix, Knuth Monika, Nikolaus Katharina, Müller Achim. Safety and accuracy of a new long-term subconjunctival glucose sensor.. Journal of diabetes. 2012. DOI: 10.1111/j.1753-0407.2012.00192.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.