Electron irradiation of conjunctival lymphoma--Monte Carlo simulation of the minute dose distribution and technique optimization.

Brualla Lorenzo, Zaragoza Francisco J, Sempau Josep, Wittig Andrea, Sauerwein Wolfgang

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: External beam radiotherapy is the only conservative curative approach for Stage I non-Hodgkin lymphomas of the conjunctiva. The target volume is geometrically complex because it includes the eyeball and lid conjunctiva. Furthermore, the target volume is adjacent to radiosensitive structures, including the lens, lacrimal glands, cornea, retina, and papilla. The radiotherapy planning and optimization requires accurate calculation of the dose in these anatomical structures that are much smaller than the structures traditionally considered in radiotherapy. Neither conventional treatment planning systems nor dosimetric measurements can reliably determine the dose distribution in these small irradiated volumes.

METHODS AND MATERIALS: The Monte Carlo simulations of a Varian Clinac 2100 C/D and human eye were performed using the penelope and penEasyLinac codes. Dose distributions and dose volume histograms were calculated for the bulbar conjunctiva, cornea, lens, retina, papilla, lacrimal gland, and anterior and posterior hemispheres.

RESULTS

The simulated results allow choosing the most adequate treatment setup configuration, which is an electron beam energy of 6 MeV with additional bolus and collimation by a cerrobend block with a central cylindrical hole of 3.0 cm diameter and central cylindrical rod of 1.0 cm diameter.

CONCLUSIONS

Monte Carlo simulation is a useful method to calculate the minute dose distribution in ocular tissue and to optimize the electron irradiation technique in highly critical structures. Using a voxelized eye phantom based on patient computed tomography images, the dose distribution can be estimated with a standard statistical uncertainty of less than 2.4% in 3 min using a computing cluster with 30 cores, which makes this planning technique clinically relevant.

Для цитирования: Brualla Lorenzo, Zaragoza Francisco J, Sempau Josep, Wittig Andrea, Sauerwein Wolfgang. Electron irradiation of conjunctival lymphoma--Monte Carlo simulation of the minute dose distribution and technique optimization.. International journal of radiation oncology, biology, physics. 2012. DOI: 10.1016/j.ijrobp.2011.09.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.