DOI 10.1016/j.jclinane.2012.05.003
Comparison of temporal artery, nasopharyngeal, and axillary temperature measurement during anesthesia in children.
Аннотация
STUDY
OBJECTIVE
To evaluate the accuracy and precision of a new, noninvasive infrared thermometer applied to the temporal artery.
DESIGN
Prospective randomized study.
SETTING
Trakya University Hospital.
PATIENTS: 60 ASA physical status 1 and 2 children undergoing surgery.
INTERVENTIONS: During anesthesia, temperature measurements were recorded with three different techniques: temporal artery, nasopharynx, and axillary temperature.
MEASUREMENTS: Temperatures measured from the nasopharynx, temporal artery, and the axilla were recorded at 15-minute intervals for the first hour, then at 30-minute intervals until the completion of surgery. During each measurement, heart rate and midarterial pressure were recorded.
MAIN
RESULTS
There were no statistically significant differences between temperatures recorded at the temporal artery and nasopharynx at 15, 30, 45, 60, 90, and 120 minutes, and the completion of surgery. Axillary temperatures were statistically lower than those recorded at the nasopharynx and the temporal artery (P < 0.001). Bland-Altman plots showed a correlation of temperature measurements between the temporal artery and nasopharyngeal methods. The axillary method had a lower correlation with the temporal artery and the nasopharyngeal methods.
CONCLUSIONS
The temporal artery thermometer is a substitute for the nasopharyngeal thermometer for core temperature measurement during anesthesia in children.