Evaluation of factors influencing resonance frequency analysis values, at insertion surgery, of implants placed in sinus-augmented and nongrafted sites.

Degidi Marco, Daprile Giuseppe, Piattelli Adriano, Carinci Francesco

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The immediate loading technique requires a high primary stability. Resonance frequency analysis (RFA) has been proposed to assess this stability with a quantitative method.

PURPOSE: The aim of the present study was to evaluate if a good primary stability could be achieved in sites that had undergone a sinus augmentation procedure and also to evaluate the importance of different clinical factors in the determination of resonance frequency values at implant insertion.

MATERIALS AND

METHODS

In 14 patients, 80 implants were inserted. Sixty-three implants were inserted in a site previously treated with a sinus augmentation procedure, while 17 implants were inserted in healed or postextraction sites. For each implant, diameter, length, bone density, insertion torque, RFA value, and percentage of implant fixed to a nongrafted bone were recorded.

RESULTS

Grafted sites showed high RFA values. A statistically significant positive correlation was found between resonance frequency values and implant diameter (p=0.007), implant length (p=0.02), diameter of the last bur used (p=0.01). No statistically significant correlation between RFA values and all the other variables considered was found.

CONCLUSIONS

Sites treated with sinus augmentation procedures can offer good primary stability after 6 months of healing. The length and diameter of the implants, together with the geometry of the implant used, are important to obtain high RFA values.

Для цитирования: Degidi Marco, Daprile Giuseppe, Piattelli Adriano, Carinci Francesco. Evaluation of factors influencing resonance frequency analysis values, at insertion surgery, of implants placed in sinus-augmented and nongrafted sites.. Clinical implant dentistry and related research. 2007. DOI: 10.1111/j.1708-8208.2007.00042.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.