DOI 10.1016/s1386-6532(09)70007-1

Медицинская библиотека · Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology · 2009

Analytical characterization of the APTIMA HPV Assay.

Dockter Janel, Schroder Astrid, Eaton Barbara, Wang Ann, Sikhamsay Nathan, Morales Liezel, Giachetti Cristina

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Human papillomavirus (HPV) testing has improved the sensitivity for the detection of cervical pre-cancer and cancer as compared to Pap testing. Several HPV tests are commercially available and most target the DNA from 13 or 14 high-risk HPV types. The APTIMA HPV Assay however, detects HPV E6/E7 mRNA from 14 high-risk types of HPV: 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66, and 68.

OBJECTIVE

To determine the analytical performance characteristics of the APTIMA HPV Assay.

STUDY

DESIGN

Analytical sensitivity, analytical specificity, reproducibility, and the effect of potentially interfering substances was determined for the APTIMA HPV Assay on both the DTS (semi-automated) and TIGRIS DTS (fully automated) systems.

RESULTS

The 95% detection limit for both systems was between 17 and 488 copies/reaction, depending on the HPV type. The assay did not cross-react with normal flora and opportunistic organisms that may be found in cervical samples, or low-risk HPV types. Spermicides, anti-fungal and anti-itch medications, whole blood, glacial acetic acid, and most lubricants did not interfere with assay performance. Those lubricants containing polyquaternium 15 did interfere with assay performance. Inter-instrument, inter-operator, inter-lot, and inter-run signal variability were <10% for >99% of the data. Intra-run variability was <15%, except for those samples with concentrations at or below the 95% detection limit of the assay.

CONCLUSIONS

Based upon the analytical sensitivity, analytical specificity, and low variability, the APTIMA HPV Assay showed excellent performance and robustness.

Для цитирования: Dockter Janel, Schroder Astrid, Eaton Barbara, Wang Ann, Sikhamsay Nathan, Morales Liezel, Giachetti Cristina. Analytical characterization of the APTIMA HPV Assay.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2009. DOI: 10.1016/s1386-6532(09)70007-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.