MR-guided HIFU treatment of symptomatic uterine fibroids using novel feedback-regulated volumetric ablation: effectiveness and clinical practice.
Аннотация
PURPOSE: To evaluate a novel feedback-regulated volumetric sonication method in MRguided HIFU treatment of symptomatic uterine fibroids.
MATERIALS AND
METHODS
27 fibroids with an average volume of 124.9 ± 139.8 cc in 18 women with symptomatic uterine fibroids were ablated using the new HIFU system Sonalleve (1.5 TMR system Achieva, Philips). 21 myomas in 13 women were reevaluated 6 months later. Standard (treatment) cells (TC) and feedback-regulated (feedback) cells (FC) with a diameter of 4, 8, 12, and 16 mm were used and compared concerning sonication success, diameter of induced necrosis, and maximum achieved temperature. The non-perfused volume ratio (NPV related to myoma volume) was quantified. The fibroid volume was measured before, 1 month, and 6 months after therapy. Symptoms were quantified using a specific questionnaire (UFS-QoL).
RESULTS
In total, 205 TC and 227 FC were applied. The NPV ratio was 23 ± 15 % (2 – 55). The TC were slightly smaller than intended (-3.9 ± 52 %; range, -100 – 81), while the FC were 20.1 ± 25.3 % bigger (p = 0.02). Feedback mechanism is less diversifying in diameter (p < 0.001). Overall, the FC correlate well with the planned treatment diameter (r = 0.79), other than the TC (r = 0.38). Six months after therapy, the fibroid volume was reduced by 45 ± 21 % (5 – 100) (p = 0.001). The symptoms decreased significantly (p = 0.001). No serious adverse events were recorded.
CONCLUSION
Use of volumetric sonication leads to homogenous heating and sufficient necrosis. It is a safe and effective therapy for treating symptomatic uterine fibroids. Successful sonication of feedback cells leads to more contiguous necrosis in diameter
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.