Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2022
Whole-body photoneutron 360° angular distribution dosimetry by novel "Sohrabi neutron dosimetry methods".
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Sohrabi Mehdi, Ebrahimzadeh Torkamani Morteza, Ali Nedaie Hassan
Аннотация
PURPOSE: Whole-body photoneutron (PN) energy-specific dosimetry on phantom surfaces and 360°angular fast PN distribution dosimetry on and around polyethylene (PE) phantom organs for prostate cancer in 10 cm × 10 cm and 20 cm × 20 cm field sizes of 18 MV X-rays of Varian Clinac 2100C linear accelerator.
METHODS: Novel "Sohrabi neutron dosimetry methods" including "miniature passive neutron dosimeter/spectrometer" and "strip polycarbonate neutron dosimeters" were applied. Energy-specific PN dose equivalents on surface as well as 360°fast PN dose equivalent angular distributions on and around PE phantom organs; head, neck, thorax, arms, pelvis, thighs and legs were determined.
RESULTS: Matrix of surface (skin) energy-specific PN dose equivalents including total thermal, total epithermal, total fast, sum total thermal + epithermal and sum total thermal + epithermal + fast, and 360°angular fast PN dose equivalent responses were determined. The results indicate that data matrix of 20 cm × 20 cm field size provides higher PN dose equivalent values than those of 10 cm × 10 cm field size for surface points even remote from the central axis and for all 360°angular fast PN dose equivalent distributions on organs studied.
CONCLUSIONS: Matrix of energy-specific PN dose equivalents was obtained demonstrating that the Sohrabi neutron dosimetry methods applied are unique for energy-specific PN dose equivalent studies as well as for 360°angular PN dose equivalent distribution data for PN-SPC risk estimation. The dosimetry methods can be specifically applied to many other exotic applications in health physics, medical physics, space flight dosimetry, and nuclear science and technology.
METHODS: Novel "Sohrabi neutron dosimetry methods" including "miniature passive neutron dosimeter/spectrometer" and "strip polycarbonate neutron dosimeters" were applied. Energy-specific PN dose equivalents on surface as well as 360°fast PN dose equivalent angular distributions on and around PE phantom organs; head, neck, thorax, arms, pelvis, thighs and legs were determined.
RESULTS: Matrix of surface (skin) energy-specific PN dose equivalents including total thermal, total epithermal, total fast, sum total thermal + epithermal and sum total thermal + epithermal + fast, and 360°angular fast PN dose equivalent responses were determined. The results indicate that data matrix of 20 cm × 20 cm field size provides higher PN dose equivalent values than those of 10 cm × 10 cm field size for surface points even remote from the central axis and for all 360°angular fast PN dose equivalent distributions on organs studied.
CONCLUSIONS: Matrix of energy-specific PN dose equivalents was obtained demonstrating that the Sohrabi neutron dosimetry methods applied are unique for energy-specific PN dose equivalent studies as well as for 360°angular PN dose equivalent distribution data for PN-SPC risk estimation. The dosimetry methods can be specifically applied to many other exotic applications in health physics, medical physics, space flight dosimetry, and nuclear science and technology.