DOI 10.1016/j.ejmp.2021.12.014

Медицинская библиотека · Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2021

Monte Carlo simulations and phantom validation of low-dose radiotherapy to the lungs using an interventional radiology C-arm fluoroscope.

Roa D, Leon S, Paucar O, Gonzales A, Schwarz B, Olguin E, Moskvin V, Alva-Sanchez M, Glassell M, Correa N, Moyses H, Shankar A, Hamrick B, Sarria G R, Li B, Tajima T, Necas A, Guzman C, Challco R, Montoya M, Meza Z, Zapata M, Gonzales A, Marquez F, Neira R, Vilca W, Mendez J, Hernandez J

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To use MC simulations and phantom measurements to investigate the dosimetry of a kilovoltage x-ray beam from an IR fluoroscope to deliver low-dose (0.3-1.0 Gy) radiotherapy to the lungs.

MATERIALS AND

METHODS

PENELOPE was used to model a 125 kV, 5.94 mm Al HVL x-ray beam produced by a fluoroscope. The model was validated through depth-dose, in-plane/cross-plane profiles and absorbed dose at 2.5-, 5.1-, 10.2- and 15.2-cm depths against the measured beam in an acrylic phantom. CT images of an anthropomorphic phantom thorax/lungs were used to simulate 0.5 Gy dose distributions for PA, AP/PA, 3-field and 4-field treatments. DVHs were generated to assess the dose to the lungs and nearby organs. Gafchromic film was used to measure doses in the phantom exposed to PA and 4-field treatments, and compared to the MC simulations.

RESULTS

Depth-dose and profile results were within 3.2% and 7.8% of the MC data uncertainty, respectively, while dose gamma analysis ranged from 0.7 to 1.0. Mean dose to the lungs were 1.1-, 0.8-, 0.9-, and 0.8- Gy for the PA, AP/PA, 3-field, and 4-field after isodose normalization to cover ∼ 95% of each lung volume. Skin dose toxicity was highest for the PA and lowest for the 4-field, and both arrangements successfully delivered the treatment on the phantom. However, the dose distribution for the PA was highly non-uniform and produced skin doses up to 4 Gy. The dose distribution for the 4-field produced a uniform 0.6 Gy dose throughout the lungs, with a maximum dose of 0.73 Gy. The average percent difference between experimental and Monte Carlo values were -0.1% (range -3% to +4%) for the PA treatment and 0.3% (range -10.3% to +15.2%) for the 4-field treatment.

CONCLUSION

A 125 kV x-ray beam from an IR fluoroscope delivered through two or more fields can deliver an effective low-dose radiotherapy treatment to the lungs. The 4-field arrangement not only provides an effective treatment, but also significant dose sparing to healthy organs, including skin, compared to the PA treatment. Use of fluoroscopy appears to be a viable alternative to megavoltage radiation therapy equipment for delivering low-dose radiotherapy to the lungs.

Для цитирования: Roa D, Leon S, Paucar O, Gonzales A, Schwarz B, Olguin E, Moskvin V, Alva-Sanchez M, Glassell M, Correa N, Moyses H, Shankar A, Hamrick B, Sarria G R, Li B, Tajima T, Necas A, Guzman C, Challco R, Montoya M, Meza Z, Zapata M, Gonzales A, Marquez F, Neira R, Vilca W, Mendez J, Hernandez J. Monte Carlo simulations and phantom validation of low-dose radiotherapy to the lungs using an interventional radiology C-arm fluoroscope.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2021. DOI: 10.1016/j.ejmp.2021.12.014.