Analysis of motor axonal damage in carpal tunnel syndrome using MScanFit MUNE.

Xu Xi, Zhan Shuqin, Wang Heying, Liu Jingjie, Xiang Li, Zhang Bo, Huang Dengchao, Gong Feng

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Using a motor unit number estimation (MUNE) method, MScanFit MUNE enables simple, fast, noninvasive, and automated analysis of the number of motor units in patients.

PURPOSE: To investigate the validity of the MScanFit MUNE method (number of motor units) against the compound muscle action potential method (compound amplitude of motor units) at identifying of motor axonal damage in carpal tunnel syndrome, particularly in the early stages, thereby providing a basis for timely treatment and surgical intervention.

STUDY

DESIGN

Electrophysiological data were collected from 170 hands of 128 patients with different stages of carpal tunnel syndrome. A further 30 healthy subjects were included. Observe the motor axonal damage at different stages.

METHODS

Compound amplitude of motor units and MScanFit MUNE values were used to analyze differences in motor axonal damage between the mild, moderate, severe patient groups, and control group. Plot the ROC curves for diagnosing carpal tunnel syndrome using the two methods.

RESULTS

The mean MScanFit MUNE values for the mild, moderate, and severe groups were 79.4 ± 19.2, 55.2 ± 17.9, and 27.4 ± 11.7, respectively. The compound amplitude of motor units for these groups were 6.5 ± 2.4 mV, 5.2 ± 2.9 mV, and 3.1 ± 1.7 mV, respectively. The values for both methods were lower in the severe group than in the mild and moderate groups (p < 0.05). The MScanFit MUNE values were lower in the mild group than in the control group (p < 0.05), whereas this difference was not observed in another method. The respective areas under the ROC curves for diagnosis were 0.78 for compound amplitude of motor units and 0.86 for MScanFit MUNE.

CONCLUSIONS

The MScanFit MUNE method is more effective in identifying motor axonal damage at different stages of carpal tunnel syndrome, especially in the early stages.

Для цитирования: Xu Xi, Zhan Shuqin, Wang Heying, Liu Jingjie, Xiang Li, Zhang Bo, Huang Dengchao, Gong Feng. Analysis of motor axonal damage in carpal tunnel syndrome using MScanFit MUNE.. Journal of hand therapy : official journal of the American Society of Hand Therapists. 2025. DOI: 10.1016/j.jht.2025.09.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.