MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.anpedi.2011.09.019

Медицинская библиотека · Anales de pediatria (Barcelona, Spain : 2003) · 2011

[Comparison between conventional ph measurement and multichannel intraluminal esophageal impedance in children with respiratory disorders].

Tolín Hernani M, Crespo Medina M, Luengo Herrero V, Martínez López C, Salcedo Posadas A, Alvarez Calatayud G, Morales Pérez J L, Sánchez Sánchez C

Сравнительное исследованиеАннотация на английскомСтатьяПеревод в очереди

Аннотация

INTRODUCTION

Gastroesophageal reflux (GER) is associated with respiratory symptoms. The link between non-acid GER and the pathogenesis of respiratory disease has been demonstrated. Esophageal multichannel intraluminal impedance (MII) is able to detect non-acid and alkaline GER, as well as reflux height. The objective of the study was to compare the diagnostic effectiveness of dual-channel pH-meter and MII.

PATIENTS AND

METHODS

A prospective study was conducted on patients diagnosed with uncontrolled asthma, persistent cough, or chronic laryngitis. Patients were monitored continuously for 24 hours using a combination of MII and a dual-channel pH-meter. A descriptive and comparative analysis of the techniques was performed using the t test for comparison between groups and McNemar test for non-parametric data.

RESULTS

A total of 49 patients with respiratory disease between September 2008 and April 2010 (79.6% uncontrolled asthma, 10.2% persistent cough, and 10.2% chronic laryngitis) were included in the study. The mean number of refluxes detected was 18.3 (range 0-93) using the pH-meter and 39.2 (11-119) using MII (P<.001). Acid GER was detected using pH in 7 children and using MII in 25 children (8 acid, 10 alkaline and 7 mixed). A mean of 21 proximal refluxes were detected using MII.

CONCLUSIONS

MII makes it possible to diagnose a greater number of refluxes, whether acid or alkaline, than conventional pH measurement in children with respiratory disease that is poorly controlled with their usual treatment. MII can also detect proximal refluxes.

Для цитирования: Tolín Hernani M, Crespo Medina M, Luengo Herrero V, Martínez López C, Salcedo Posadas A, Alvarez Calatayud G, Morales Pérez J L, Sánchez Sánchez C. [Comparison between conventional ph measurement and multichannel intraluminal esophageal impedance in children with respiratory disorders].. Anales de pediatria (Barcelona, Spain : 2003). 2011. DOI: 10.1016/j.anpedi.2011.09.019.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.